Case report
Artefact or hip prosthetic fracture on multislice CT? The importance of
correct positioning when scanning metal implants
M.W. Kusk
Hospital of Southwest Jutland, Department of Radiology and Nuclear Medicine, Finsensgade 35, 6700 Esbjerg, Denmark
article info
Article history:
Received 27 July 2018
Received in revised form
27 November 2018
Accepted 3 December 2018
Available online 14 December 2018
Introduction
Fractures of total hip arthroplasty (THA) implants, are rare but not
unknown.
1
Correct diagnosis is crucial in deciding whether implant
replacement surgery is indicated. Loosening of femoral stem
or acetabular components are far more common,
2
but cannot be
differentiated without imaging. Conventional radiography of the hip
is first-line modality, but when the diagnosis is unclear or incon-
sistent with clinical presentation, CT is the modality of choice.
3,4
However metal implants impart artefacts on CT images. Beam
hardening artefacts occur where x-rays are filtered by passage
through radiodense material, increasing average energy of the
x-ray spectrum.
5
The result is a dark streak- or star-like artefact
radiating from the metal into adjacent tissue, and potentially
obscuring pathology. Artefact reduction can be achieved by
increasing kVp, metal artefact reduction (MAR) reconstruction al-
gorithms, and reducing thickness of metal in the scanplane,
through careful positioning of patient and/or tilting of the scanner
gantry.
6
When these precautions are not observed, there is a risk of
the artefacts simulating pathology.
Case report
A 67 year old man presented at the orthopedic surgery depart-
ment 6 months after right THA, with increasing pain when walking,
and no history of hip trauma. Paraclinical findings did not indicate
infection. Standard hip radiographs were performed, showing
normal postoperative status. The patient was then referred to CT of
the hip prosthesis, to rule out component loosening. Scanning was
on a Toshiba Aquilion PRIME 80-slice CT-scanner (Toshiba Medical
Systems, The Netherlands) with a dedicated protocol as follows:
mA: 300, kVp 135, Pitch: 0,684, rotation time: 1 sec. Images were
reconstructed in 0,5 mm thickness and 0,3 mm interval at a medium
sharp kernel, for MPR processing. Images were reconstructed with
and without Single Energy Metal Artefact Reduction (SEMAR).
Images were then reviewed by a musculoskeletal (MSK) radi-
ologist with more than 15 years of experience. Small cysts were
found in the acetabulum/liner interface. At the neck of the femoral
component, a dark streak raised suspicion of fractured prosthesis
(Fig. 1), on both SEMAR and non-SEMAR images, especially
on paracoronal MPR. Volume Rendering Technique (VRT) re-
constructions were used for a global view. The irregular appearance
and localization could indicate serrated edges of a material fracture.
The case was discussed with another MSK radiologist. A senior
CT radiographer was consulted to see if artefacts could be resolved
by reconstructions at different algorithms, which was not the case.
Therefore the patient was recalled for renewed CT scan of the re-
gion, with identical parameters. This time the patient was posi-
tioned with the scan plane as perpendicular to the femoral neck as
possible, reducing the amount of metal traversed by the beam.
The patient was positioned as shown in Fig. 2, as THA patients are
advised not to cross their legs (to minimize subluxation risk). A
board was placed under the shoulder region, and feet immobilized
in a vacuum bag, taped to the patient table but extending beyond it.
The dark streak could not be reproduced, and it was concluded that
no fracture was present (Fig. 3).
Discussion
This case highlights the importance of correct patient posi-
tioning in CT, when scanning metal implants, even though
advanced MAR techniques are used. MAR algorithms are designed
to reduce artefacts in tissue surrounding the metal, which may help
to explain the failure to reduce the artefact within the metal.
SEMAR works by identifying projections in raw data space
containing metal artefacts, subtracting these from the sinogram
and interpolating missing projections from remaining data.
Following this, new images are reconstructed from the resulting
sinogram, with the metal data reinserted. Other vendors offer
similar solutions: MAR (GE), O-MAR (Philips), I-MAR (Siemens).
7
E-mail address: martin.weber.kusk@rsyd.dk.
Contents lists available at ScienceDirect
Radiography
journal homepage: www.elsevier.com/locate/radi
https://doi.org/10.1016/j.radi.2018.12.001
1078-8174/© 2018 The College of Radiographers. Published by Elsevier Ltd. All rights reserved.
Radiography 25 (2019) e75ee77