Ultramicroscopy 8 (1982) 137 144 137
North-Holland Publishing Company
RESOLUTION IN SURFACE SCANNING ELECTRON MICROSCOPY OF BULK SAMPLES
A. N. Broers
IBM, East Fishkill, Route 52, Hopewell Junction, N. Y.
Ultimately resolution in today's surface scanning electron microscopes is determined by electron-sample
interactions and not by the capabilities of the electron optical systems. In other words, it is not possible to
resolve surface features on bulk samples (i.e., samples that are opaque to electrons) that are separated by as small
a distance as might be expected from the minimum beam diameter.
Tl~e minimum beam diameter in a standard secondary electron SEM is set by the spherical aberration of the final
lens and diffraction to less than 2.0nm, but secondary electrons are excited up to several nanometers away from
the incident electrons. As a result, the best point to point resolution for secondary electron images of features on
the surface of true bulk samples has been 5nm - 10nm.
High energy electrons elastically scattered from a very shallow surface layer are also used for imaging the surface
of bulk samples. A shorter focal length final lens is used in this case because the sample can be immersed in the
magnetic field. This allows the beam diameter to be reduced to 0.5nm but again interaction phenomena limit
resolution. Finite penetration of the electrons before they are scattered apparently limits point to point resolution
to about 2nm. Contrast is high,however, and ultimately it may be possible to "see" single atomic surface steps.
The secondary electron surface scanning electron microscope electron surface image provides higher resolution and contrast
(SEM) 1 has become almost as ubiquitous as the optical micro-
scope. SEM images are easy to interpret and sample prepara-
tion, particularly for electrically conducting samples, is very
simple. The SEM provides higher resolution and greater depth
of field than the optical microscope, and a variety of special
operating modes provide information about electrical and
magnetic properties of the sample that cannot be obtained
optically.
SEM resolution is much higher than that of the optical micro-
scope, but it is not as high as that of the TEM (Transmission
Electron Microscope) or of the STEM (Scanning Transmission
Electron Microscope). There are two major reasons for this.
First, in order to collect secondary electrons efficiently, the
sample has to be placed outside the magnetic field of the final
lens. This means the microscope must have a final lens with a
longer focal length than the lenses used in transmission instru-
ments where the thin sample can be immersed in the lens field.
Longer focal length means higher aberrations and these aberra-
tions, combined with the lower accelerating voltage generally
used in SEM's, lead to a minimum beam size up to six times
larger. Lower accelerating voltage is used to prevent the elec-
tron beam from penetrating completely through surface protru-
sions, thus producing secondary electrons from both sides of
the protrusions and "flairing" out these regions of the image.
The larger beam size is not as important as it might seem,
however, because the second problem encountered with the
secondary electron surface image is that the size of the area
from which secondary electrons are emitted is even larger than
the beam. The best point to point resolution measured in a
secondary electron SEM is about 5rim, although, as we shall
see, the beam diameter can be 2nm.
In attempts to avoid these restrictions, methods have been
developed that allow surface images to be formed with primary
electrons scattered from the sample. In this case a short focal
length lens can be used because the sample can be immersed in
the lens field. This means the beam can be much smaller, but
again electron interaction phenomena apparently limit resolu-
tion. The distance below the surface from which electrons are
scattered is such that the area of the surface that affects the
image is again larger than the beam. Nonetheless, point to
point resolution is reduced to about 2nm, The high energy
than the secondary electron image, but is generally not as
useful because the size of the sample is constrained by the
necd to fit it between the lens pole-pieces, and the sample has
to be examined at a relatively steep angle. (>45 ° from nor-
mal).
There is the possibility for a method that completely avoids
"interaction volume" limitations. This is to form a scanning
image similar to the glancing angle projection image reported
by Yagi 2. Here contrast arises from interference between
waves excited from the top and bottom of surface steps, and
does not depend on the types of excitation phenomena used in
the other surface images. This phase-contrast image is restrict-
ed to smooth samples, and to viewing at very steep angles.
The angle of incidence (88 ° from normal) is so steep that the
image is virtually one dimensional, but resolution approaching
atomic scale can be obtained. Cowley has investigated this
type of image. 3
Electron Optics
The diameter of the electron beam in an SEM depends on the
accelerating voltage (V(v)), the normalized brightness
J3(A/cm2.steradian.V) and axial energy spread 6V(v) of the
electron gun, and on the aberrations of the final lens. With
electron guns employing field emission or lanthanum hexabor-
ide cathodes it is possible to closely approach the theoretical
minimum (dlim) set by the final lens spherical aberration coef-
ficient Cs(m) and diffraction and still retain adequate current
to form an image with acceptable signal to noise ratio (S/N).
dlinl = 0.9 Csl/4 )~3/4 (m)
- wavelength of the electrons (m)
This relationship assumes that chromatic aberration and astig-
matism are negligible, For SEM's that use thermal cathodes
this is not valid, because Boersch effect 4 typically increases
chromatic spread to the point that chromatic aberration is
more important than spherical aberration. 5
The degree to which dli m is approached by guns of different
brightness and energy spread can be seen in fig. 1. Fig. 1
0304-3991/82/0000 0000/$02.75 © 1982 North-Holland