Accuracy Improvements in LPC Measurements for CMP Slurries
Bruno Tolla and David Boldridge
Cabot Microelectronics Corporation, 870 Commons Drive, Aurora, IL 60504, USA
ABSTRACT
We have examined the Large Particle Count (LPC) analytical method to see whether
there are opportunities to improve both the accuracy and precision in hope of improving the
utility of the LPC measurement. We have identified weaknesses in the current method that limit
both its accuracy and its precision, and which can introduce count errors in excess of a factor of
10. We propose modifications to the current method which result in both accuracy and precision
improvements. We recommend these improvements as absolutely necessary for any experiments
designed to test the correlation between LPC and defectivity.
INTRODUCTION
Chemical Mechanical Planarization depends on specially designed slurries to produce a
flat, defect free surface on a semiconductor wafer in preparation for subsequent processing
steps.[1,2] These slurries consist of abrasive particles and active chemistry in a liquid carrier,
typically water. To assure effective polishing, the abrasive particle size distribution is tightly
controlled. The practitioners of CMP are understandably worried that small quantities of
undesirably large abrasive or impurity particles could damage the wafers and lead to yield loss.
The quality metric of Large Particle Count (LPC) was implemented to help guard against
abnormally high levels of these large particles.
The LPC is presumed to indicate the number of particles larger than a given size, usually
stated as 0.56 m or 1.01 m. These particles represent only ppm of the mass of the slurry and
ppb of the number of abrasive particles, necessitating the use of a highly selective measurement
technology. The task of quantifying such a small fraction of the total particle size distribution is
quite challenging, and the industry has implemented Single Particle Optical Sizing (SPOS) as a
routine analytical technique.[3-8] This technique is nominally capable of detecting only the
largest particles while ignoring the smaller, necessary abrasive component. While there is ample
evidence that the SPOS technique can provide a warning signal in extreme cases, continued
improvements in abrasive and slurry production have dramatically reduced the typical LPC
levels. As a result, the correlation between LPC and defectivity has become much less
clear.[3,9-13]
Weak correlation between LPC and defectivity can be the result of either inadequacy of
the LPC measurement protocol or the appearance of non-LPC related defect mechanisms. We
have found compelling evidence that inappropriate calibration and operating practices can
severely compromise both the accuracy and the precision of the technique. Bringing the
calibration and operating procedures into conformance with best known practice are required if
the test is have any hope of recovering some of its former utility.
Mater. Res. Soc. Symp. Proc. Vol. 1157 © 2009 Materials Research Society 1157-E04-07