Use of radiochromic film as a high-spatial resolution dosimeter
by Raman spectroscopy
Jamal Ahmad Mirza and Hyeonsuk Park
Program in Biomedical Radiation Sciences, Department of Transdisciplinary Studies, Graduate School
of Convergence Science and Technology, Seoul National University, Seoul 08826, South Korea
So-Yeon Park
Interdisciplinary Program in Radiation Applied Life Sciences, Seoul National University College of Medicine,
Seoul 03080, South Korea
Sung-Joon Ye
a)
Program in Biomedical Radiation Sciences, Department of Transdisciplinary Studies, Graduate School
of Convergence Science and Technology, Seoul National University, Seoul 08826, South Korea;
Interdisciplinary Program in Radiation Applied Life Sciences, Seoul National University College of Medicine,
Seoul 03080, South Korea; and Center for Convergence Research on Robotics, Advance Institutes
of Convergence Technology, Seoul National University, Suwon 16229, South Korea
(Received 1 March 2016; revised 14 June 2016; accepted for publication 20 June 2016;
published 14 July 2016)
Purpose: Due to increasing demand for high-spatial resolution dosimetry, radiochromic films have
been investigated as potential candidates but are often limited by the scanning system, e.g., flatbed
optical scanner. In this study, Raman spectroscopy in conjunction with a microscope was selected as
an alternative method for high-spatial resolution dosimetry of radiochromic film.
Methods: Unlaminated Gafchromic™ EBT3 films were irradiated with doses between 0 and 50 Gy
using 6 MV x-rays of a clinical linear accelerator. Depth profiling from the surface of unlaminated
film was performed to acquire the maximum Raman intensity peaks of C≡≡C and C==C stretching
bands of diacetylene polymer. The Raman mapping technique for a region of interest (200 × 200,
30 × 30 μm
2
) was developed to reduce a large variation in a Raman spectrum produced with a
sampling resolution of a few μm. The preprocessing of Raman spectra was carried out to determine
a dosimetric relationship with the amount of diacetylene polymerization.
Results: Due to partial diacetylene polymerization upon irradiation, two Raman peaks of C==C and
C≡≡C stretching bands were observed around 1447 and 2060 cm
−1
, respectively. The maximum
intensities of the two peaks were obtained by positioning a focused laser spot on the surface of
unlaminated film. For the dose range of 0–50 Gy, the band heights of both C≡≡C and C==C peaks
increase asymptotically with increasing doses and can be fit with an exponential function of two
components. The relative standard deviation in Raman mapping was found to be less than ±5%. By
using this technique, dose uniformity was found to be within ±2%.
Conclusions: The Raman intensity for C==C and C≡≡C peaks increases with an increase in the
amount of diacetylene polymerization due to an increase in dose. This study shows the potential
of Raman spectroscopy as an alternative for absolute dosimetry verifications with a high-spatial reso-
lution of a few μm, but these findings need to be further validated for the purpose of microdosimetry.
C
2016 American Association of Physicists in Medicine. [http://dx.doi.org/10.1118/1.4955119]
Key words: Raman spectroscopy, high-spatial resolution, microdosimetry, radiochromic film
1. INTRODUCTION
Recent technological progress in radiological imaging and
therapy instruments has greatly improved tumor localization
and management. Such innovative instruments often require
dosimetry systems to fulfill extraordinary treatment planning
and in vivo needs, e.g., evaluating the dose in regions with
high gradients, at buildup and interface regions, regions in
close proximity to brachytherapy sources, and for small
field dosimetry.
1,2
Conventional radiation detectors, like an
ionization chamber or a semiconductor detector, have certain
limitations, such as a relatively large sensitive volume or
nonwater equivalence. Other unsuitable dosimeters include
thermoluminescent detectors due to the time-consuming
readout procedure and inability to archive their dosimetric
data, and radiographic films that are nonwater equivalent,
sensitive to light, and require wet chemical processing.
3
On
the other hand, polymer gels and radiochromic films have
been considered suitable for these clinical needs because of
their high-spatial resolution and near water equivalence.
1
For
radiobiological considerations, a very high-spatial resolution
dosimeter, which is capable of giving information on spatial,
temporal, and spectral distributions of energy imparted
in cellular and subcellular structures, and the associated
biological effects, is required.
4
Novel dosimeters have been
developed for the space industry, such as microelectronic
4520 Med. Phys. 43 (8), August 2016 0094-2405/2016/43(8)/4520/9/$30.00 © 2016 Am. Assoc. Phys. Med. 4520