Performance evaluation of the Abbott CELL-DYN Emerald for use
as a bench-top analyzer in a research setting
T.-L. KHOO*
, † , ‡
, N. XIROS
‡
, F. GUAN*
†
, D. ORELLANA
‡
, J. HOLST*
†
, D. E. JOSHUA
† , ‡
, J. E. J. RASKO*
,†,‡,§
*Gene and Stem Cell Therapy
Program, Centenary Institute,
Camperdown, NSW, Australia
†
Sydney Medical School,
University of Sydney, Sydney,
NSW, Australia
‡
Institute of Haematology, Royal
Prince Alfred Hospital,
Camperdown, NSW, Australia
§
Cell and Molecular Therapies,
Royal Prince Alfred Hospital,
Camperdown, NSW, Australia
Correspondence:
John E. J. Rasko, Locked Bag 6,
Newtown, NSW 2042, Australia.
Tel.: +61 2 9565 6156;
Fax: +61 2 9565 6101;
E-mail: j.rasko@centenary.org.au.
doi:10.1111/ijlh.12040
Received 31 July 2012; accepted
for publication 17 October 2012
Keywords
Analytical performance,
automated hematology
analyzer, CELL-DYN Emerald,
human full blood count, mouse
full blood count
SUMMARY
Introduction: The CELL-DYN Emerald is a compact bench-top hema-
tology analyzer that can be used for a three-part white cell differ-
ential analysis. To determine its utility for analysis of human and
mouse samples, we evaluated this machine against the larger
CELL-DYN Sapphire and Sysmex XT2000iV hematology analyzers.
Methods: 120 human (normal and abnormal) and 30 mouse (nor-
mal and abnormal) samples were analyzed on both the CELL-DYN
Emerald and CELL-DYN Sapphire or Sysmex XT2000iV analyzers.
For mouse samples, the CELL-DYN Emerald analyzer required
manual recalibration based on the histogram populations.
Results: Analysis of the CELL-DYN Emerald showed excellent preci-
sion, within accepted ranges (white cell count CV% = 2.09%;
hemoglobin CV% = 1.68%; platelets CV% = 4.13%). Linearity was
excellent (R
2
0.99), carryover was minimal (<1%), and overall
interinstrument agreement was acceptable for both human and
mouse samples. Comparison between the CELL-DYN Emerald and
Sapphire analyzers for human samples or Sysmex XT2000iV
analyzer for mouse samples showed excellent correlation for all
parameters.
Conclusion: The CELL-DYN Emerald was generally comparable to
the larger reference analyzer for both human and mouse samples.
It would be suitable for use in satellite research laboratories or as a
backup system in larger laboratories.
INTRODUCTION
Automated hematology analyzers have become inte-
gral in modern hematology laboratories, with their
ability to handle a large volume of specimens
efficiently. They generally exhibit good accuracy,
precision, and a fast turnaround time. Relatively few
performance differences between instruments are
observed in the analysis of ‘normal’ samples [1], but
variations become more pronounced when there are
‘abnormal’ populations, for example immature granu-
locytes and nucleated red blood cells. However, the
complexity, size, and cost of large instruments may
limit their use in a point of care setting, such as in the
emergency department, intensive care units, and out-
patient clinics that do not require a high-throughput
© 2012 John Wiley & Sons Ltd, Int. Jnl. Lab. Hem. 2013, 35, 447–456 447
ORIGINAL ARTICLE INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY
International Journal of Laboratory Hematology
The Official journal of the International Society for Laboratory Hematology