Analytica Chimica Acta 418 (2000) 197–203
A bromide ion-selective polymeric membrane electrode based
on a benzo-derivative xanthenium bromide salt
Mojtaba Shamsipur
a,∗
, Shohreh Rouhani
b
, Ali Mohajeri
c
,
Mohammad Reza Ganjali
c
, Parviz Rashidi-Ranjbar
c
a
Department of Chemistry, Razi University, Kermanshah, Iran
b
Department of Chemistry, Bou-Ali Sina University, Hamadan, Iran
c
Department of Chemistry, Tehran University, Tehran, Iran
Received 29 December 1999; received in revised form 17 April 2000; accepted 1 May 2000
Abstract
A novel bromide ion-selective PVC membrane sensor based on 14-phenyldibenzo[a,j]xanthenium bromide as a membrane
carrier was prepared. The electrode has a linear dynamic range between 1.0×10
−1
and 3.2×10
−5
M with a near-Nernstian
slope of 61±1 mV per decade and a detection limit of 2.0×10
−5
M. The potentiometric response is independent of the pH
of the solution in the pH range 4.5–8.5. The electrode possesses the advantages of low resistance, fast response and good
selectivities over a variety of other anions. It was applied as an indicator electrode in potentiometric titration of bromide ions.
© 2000 Elsevier Science B.V. All rights reserved.
Keywords: Bromide ion-selective electrode; 14-Phenyldibenzo[a,j]xanthenium bromide; PVC membrane; Potentiometry
1. Introduction
To date, a wide variety of carrier-based ion-selective
polymeric membranes for chloride [1–8] and iodide
[9–15] ions have been reported in the literature and
their widespread use in clinical chemistry are well
established [16–19]. These electrodes usually employ
various types of anion carriers including lipophilic
cationic salts, quaternary ammoniums, organometal-
lic compounds, vitamin B
12
derivatives, metallopor-
phyrins, Schiff base complexes and other cationic
complexes. In comparison, the literature reports on
the ion-selective sensors that can directly monitor
Br
−
ions in different aqueous solutions are quite
∗
Corresponding author. Tel.: +98-831-723307;
fax: 98-831-831618.
sparse [20–22]. The reported bromide ion-selective
electrodes are based on heterogeneous membranes
containing silver bromide. To the best of our knowl-
edge, there is no previous report on the carrier-based
ion-selective membrane sensors for bromide ion.
We have recently introduced PVC-based membrane
sensors for K
+
[23], Be
2+
[24], Sr
2+
[25], Zn
2+
[26],
Pb
2+
[27,28], Hg
2+
[29,30], Ag
+
[31], I
3
−
[32], I
−
[33] and SCN
−
[34] ions based on different noncyclic
and macrocyclic ligands (for cations) and metallopor-
phyrin and charge-transfer complexes (for anions).
Due to the vital importance of bromide determination
in chemical, industrial and clinical analyses [35], we
were interested in the preparation of a new solvent
polymeric membrane sensor for selective monitoring
of Br
−
ion in solution. In this paper, we report a se-
lective PVC membrane electrode for Br
−
ion based
0003-2670/00/$ – see front matter © 2000 Elsevier Science B.V. All rights reserved.
PII:S0003-2670(00)00954-5