f
Two-electrode biopotential amplifier
with current-driven inputs
D. Dobrev I. Daskalov
Centre of Biomedical Engineering, Bulgarian Academy of Sciences, Sofia, Bulgaria
Abstract--A circuit was developed for a differentia/ two-electrode biopotential
amplifier. Current sources at the amplifier inputs were controlled by the common-
mode voltage. This principle is well known in telephony for interfacing the telephone
line with analogue-type phones. A low impedance of about 1 k,Q was obtained
between each input and the common point of the circuit. The differential input
impedance of 60 Mr2 was obtained with the use of precision resistors. Considerable
reduction in the common-mode voltages of more than 200 times resulted. The circuit
can be useful for biosignal acquisition from subjects in areas of very high electro-
magnetic fields, where high common-mode voltages could saturate the input
amplifier stages.
Keywords--Amplifier, Bio-electric amplifier, Differential amplifier, Electromagnetic
interference
\
Med. Biol. Eng. Comput., 2002, 40, 122-127
J
1 Introduction
MANY EXPERIMENTAL and clinical applications connected with
biopotential measurement could benefit from the use of only two
electrodes, provided adequate signal acquisition would be
obtained. Electrocardiogram monitoring in intensive care
wards, ambulatory monitors, defibrillators, etc. are among the
most obvious examples.
One of the main problems in two- and three-electrode
differential amplifiers is the transformation of the common-
mode interference voltage into a differential signal, owing to
disturbed symmetry of the body-amplifier interface (THAKOR
and WEBSTER, 1980; WINTER and WEBSTER, 1983; PALLAS-
ARENY, 1986; METTINGVAN RIJN et al., 1990). Here, the need
to use screened patient cables can be added (WOOD et al., 1995).
Other problems can arise in connection with electrostatic
potentials, electrode polarisation voltages, electromagnetic
interference etc.
Modern biopotential amplifiers are highly isolated (floating),
as required by regulations and standards for patient safety. Thus
the conditions of interference rejection change and require
special attention (METTING VAN RIJN et al., 1991).
Another aspect of modern instrumentation is the analogue-to-
digital conversion of the biosignals, which allows the application
of efficient algorithms for power-line interference suppression.
Thus the problem of reducing power-line noise is less critical or
even practically eliminated (DASKALOV et al., 1998).
However, high-intensity power-line or other types of
common-mode voltage could become a more important
impeding factor, leading to saturation of the amplifier input
stage. Another practical aspect is the need to use non-screened
Correspondence should be addressed to Dr I. K. Daskalov;
emaih ikdas@argo.bas.bg
Paper received 3 August 2001 and in final form 30 October 2001
MBEC online number: 20023643
© IFMBE: 2002
122
wires in some applications, which can also result in an increased
common-mode signal and thus create a risk of amplifier
saturation.
These considerations stimulated us to try and design an
amplifier with low impedance of both inputs with respect to
the common point, but with adequately high differential input
impedance. The amplifier is considered here in the case of
electrocardiogram acquisition.
2 Amplifier circuit
The patient-amplifier interface circuit, with and without
isolation, and using three or two electrodes, has been investi-
gated by many authors (HUHTA and WEBSTER, 1973; THAKOR
and WEBSTER, 1980; PALLAS-ARENY, 1988; WOOD et al.,
1995). Therefore the same type of equivalent circuit is used
here, and similar designations of the respective quantities are
adopted, its well-known configuration is shown in Fig. 1.
The following impedances are considered: from power line to
patient body = Zp; from body to ground = Zb; skin-electrode =
Zel -Ze3 ; from power line to amplifier inputs - Zs; differential
amplifier input - Zj; from amplifier inputs to common floating
point = Zc; from floating common point to ground = Zg. The
power-line voltage Vpl = 220 V/50 Hz; a and b are the amplifier
inputs, and r is the amplifier reference voltage point.
The principle of the proposed circuit is shown in Fig. 2. The
differential amplifier inputs are connected to the reference point
using two current generators, driven by the common-mode
voltage from the differential pair output. Thus low input-to-
reference impedances are obtained without a reduction in the
differential input impedance. This principle is known in com-
munication engineering, where an impedance-balanced
subscriber line helps to reduce noise, and low impedance to
earth improves safety (e.g. HARDY, 1986). it was (and still is)
used to interface a two-wire telephone line to analogue tele-
phones. The circuit is called a subscriber line interface circuit
(SLIC) (e.g. LEGERITY INC., 1999).
Medical & Biological Engineering & Computing 2002, Vol. 40