25 ORIGINAL ARTICLE Sociedade Brasileira para o Estudo da Dor c Rev Dor. São Paulo, 2012 jan-mar;13(1):25-9 Effects of ΔF delivery ramp variation on the accommodation of the interferential current in healthy women* Efeitos da variação da rampa de entrega do ΔF sobre a acomodação da corrente inter- ferencial em mulheres saudáveis Thais Eduarda Carvalho Guerra 1 , Gladson Ricardo Flor Bertolini 2 * Received from the Laboratory of Injuries and Physical Therapy Resources, State University of Western Paraná (UNIOESTE). Cascavel, PR. 1. Graduated in Physical Therapy, State University of Western Paraná (UNIOESTE). Cascavel, PR, Brazil. 2. Professor of the Physical Therapy Course, State University of Western Paraná (UNIOESTE). Cascavel, PR, Brazil. Correspondence to: Gladson Ricardo Flor Bertolini Rua Universitária, 2069 – Jardim Universitário Colegiado de Fisioterapia, CCBS, Campus Cascavel da UNIOESTE 85819-110 Cascavel, PR. E-mail: gladson_ricardo@yahoo.com.br SUMMARY BACKGROUND AND OBJECTIVES: Interferen- tial current is a common electrotherapy modality in physical therapy, but the regulation of its par- ameters is still empirical, being necessary studies to establish adequate parameters for its application. This study aimed at evaluating onset time of the first accommodation and the number of times it oc- curs during 10 minutes of interferential current ap- plication, varying its presentation form of delta F (ΔF) delivery ramps. METHOD: The sample consisted of 18 women div- ided in 3 groups: Group A – 1:1 (variation every 1 second), in the 2 nd day 1:5:1 (frequencies increase and decrease in 1 second and maintenance for 5 seconds), 3 rd day – 6:6 (frequency increase and decrease in 6 seconds); Group B – 1:5:1 in the 1 st day, in the 2 nd day we used 6:6 and in the 3 rd day 1:1; Group C – 6:6 in the 1 st day, 1:1 in the 2 nd and 1:5:1 in the 3 rd day. Stimulation was above the sensory threshold during 10 minutes and volunteers would indicate the mo- ment of the irst accommodation and how many times would accommodation occur. RESULTS: During accommodation threshold evalua- tion, no ΔF delivery ramp presented signiicant differ- ence. There has been signiicant difference in total num- ber of accommodations when comparing ramp 1:5:1 to ramp 6:6 with the latter presenting the highest number of accommodations. CONCLUSION: ΔF delivery ramp has not inluenced accommodation threshold and ramp 1:5:1 had the lowest number of accommodations as compared to ramp 6:6. Keywords: Analgesia, Electric stimulation therapy, Electricity. RESUMO JUSTIFICATIVA E OBJETIVOS: A corrente interfer- encial é modalidade de eletroterapia comum na prática isioterapêutica, mas, a regulagem de seus parâmetros ainda é empírica, sendo necessários estudos que esta- beleçam os parâmetros adequados para seu uso. O objet- ivo deste estudo foi avaliar o tempo de início da primeira acomodação e o número de vezes em que ela ocorreu du- rante 10 minutos de aplicação da corrente interferencial, variando a forma de apresentação das rampas de entrega do delta F (ΔF). MÉTODO: A amostra foi composta por 18 mulheres, que foram divididas em 3 grupos: Grupo A – 1:1 (varia- ção a cada 1 segundo), no 2º dia 1:5:1 (subida e des- cida das frequências em 1 segundo e manutenção por 5 segundos), 3º dia – 6:6 (subida e descida das frequência em 6 segundos); Grupo B –1:5:1 no 1º dia, no 2º dia utilizou 6:6, e no 3º dia 1:1. Grupo C – 6:6 no 1º dia, 1:1 no 2º e 1:5:1 no 3º dia. A estimulação era acima do limiar sensitivo, durante 10 minutos, e as voluntárias indica- vam o momento em que ocorria a primeira acomodação e quantas vezes ocorria. RESULTADOS: Ao avaliar o limiar de acomodação, nenhuma das rampas de entrega de ΔF apresentou dif- erença signiicativa. Para o número total de acomoda- ções houve diferença signiicativa ao comparar a rampa