Digital Breast Tomosynthesis with Minimal Breast Compression
David A. Scaduto
*
, Min Yang, Jennifer Ripton-Snyder, Paul R. Fisher, Wei Zhao
Department of Radiology, Stony Brook University,
L4-120 Health Sciences Center, Stony Brook, New York 11794-8460
ABSTRACT
Breast compression is utilized in mammography to improve image quality and reduce radiation dose. Lesion conspicuity
is improved by reducing scatter effects on contrast and by reducing the superposition of tissue structures. However, patient
discomfort due to breast compression has been cited as a potential cause of noncompliance with recommended screening
practices. Further, compression may also occlude blood flow in the breast, complicating imaging with intravenous contrast
agents and preventing accurate quantification of contrast enhancement and kinetics. Previous studies have investigated
reducing breast compression in planar mammography and digital breast tomosynthesis (DBT), though this typically comes
at the expense of degradation in image quality or increase in mean glandular dose (MGD). We propose to optimize the
image acquisition technique for reduced compression in DBT without compromising image quality or increasing MGD.
A zero-frequency signal-difference-to-noise ratio model is employed to investigate the relationship between tube potential,
SDNR and MGD. Phantom and patient images are acquired on a prototype DBT system using the optimized imaging
parameters and are assessed for image quality and lesion conspicuity. A preliminary assessment of patient motion during
DBT with minimal compression is presented.
Keywords: digital breast tomosynthesis, breast imaging, breast compression, scatter, image quality, x-ray imaging, mean
glandular dose
1. INTRODUCTION
Mammography employs compression of the breast to both improve image quality and reduce radiation dose. Contrast
resolution is improved through breast compression by both decreasing the scatter-to-primary ratio, and utilizing lower x-
ray energy for better soft tissue contrast. Lesion conspicuity is increased by redistributing breast tissue and reducing the
superposition of structures. Additionally, compression immobilizes the breast, reducing motion artifacts during
acquisition. Finally, the decrease in breast thickness results in lower mean glandular dose (MGD).
However, patients often report discomfort associated with this compression, and this discomfort has been cited as a
potential cause for noncompliance with recommended screening practices.
1,2
Additionally, breast compression has been
shown to occlude blood flow in the breast
3,4
, a potential complication when imaging with intravenous contrast agents,
especially when contrast kinetics and enhancement quantification are investigated. Restricted blood flow due to breast
compression has been cited as a potential cause of the inability to consistently correlate contrast kinetic curves with
histopathology in some contrast-enhanced digital mammography studies
5–7
, as has been achieved in compression-free
contrast-imaging modalities like breast MRI.
8,9
Reducing the compression used in x-ray imaging of the breast is thus
desirable to both increase patient compliance and potentially improve cancer diagnosis in contrast-enhanced imaging.
Previous studies investigating the effect of breast thickness in planar mammography have shown that the consequent
increased breast thickness results in a loss of spatial
10
and contrast resolution
11
due to geometrical unsharpness and
increased scatter radiation, respectively, confirming that reducing breast compression will lead to degradation in image
quality.
Digital breast tomosynthesis (DBT) is a three-dimensional (3D) imaging technique wherein projection images acquired
around the breast are reconstructed into a quasi-3D image volume. Due to the improved tissue separation afforded in the
depth direction, DBT may not suffer the same degradation in lesion conspicuity as full-field digital mammography (FFDM)
when breast compression is reduced. However, due to the oblique angles at which projections are acquired, anti-scatter
*
david.scaduto@stonybrookmedicine.edu; phone +1-631-444-2004
Medical Imaging 2015: Physics of Medical Imaging, edited by Christoph Hoeschen, Despina Kontos,
Proc. of SPIE Vol. 9412, 94121Y · © 2015 SPIE · CCC code: 1605-7422/15/$18 · doi: 10.1117/12.2081543
Proc. of SPIE Vol. 9412 94121Y-1