Journal of Chromatography A, 1217 (2010) 683–688
Contents lists available at ScienceDirect
Journal of Chromatography A
journal homepage: www.elsevier.com/locate/chroma
Efficiency of the same neat silica column in hydrophilic interaction
chromatography and per aqueous liquid chromatography
Fabrice Gritti
a
, Alberto dos Santos Pereira
b
, Pat Sandra
b
, Georges Guiochon
a,∗
a
Department of Chemistry, University of Tennessee, Knoxville, TN 37996-1600, USA
b
Research Institute for Chromatography, Kennedypark 26, 8500 Kortrijk, Belgium
article info
Article history:
Received 13 July 2009
Received in revised form
20 November 2009
Accepted 2 December 2009
Keywords:
Hydrophilic interaction
HILIC
Hydrophobic interactions
Per aqueous liquid chromatography
PALC
Mass transfer mechanism
Halo HILIC
Caffeine
abstract
The dependencies on the mobile phase flow velocity of the efficiency of a column packed with shell
particles of neat porous silica (Halo) was measured under two different sets of experimental condi-
tions. These conditions corresponded to the retention mechanisms of per aqueous liquid chromatography
(PALC) at low acetonitrile concentrations and of hydrophilic interaction chromatography (HILIC) at high
acetonitrile concentrations. The results are compared. Small amounts of a diluted solution of caffeine
were injected in order to record the chromatograms under strictly linear conditions. These efficiencies
were measured in both water-rich (PALC retention mechanism) and acetonitrile-rich (HILIC mechanism)
mobile phases for the same retention factors, between 0.25 and 2.5. The mobile phases were mixtures of
acetonitrile and water containing neither supporting salt nor buffer component. At low retention factors,
the efficiency of caffeine is better in the PALC than in the HILIC mode. For k
′
= 0.5, the minimum reduced
height equivalent to a theoretical plate (HETP) is close to 2.5 in PALC while it exceeds 5 in HILIC. The
converse is true for high retention factors. For k
′
> 2.5, the HETP is lower in HILIC than in PALC, because
the major contribution to band broadening and peak tailing in this latter mode originates from the het-
erogeneous thermodynamics of retention and eventually restricts column performance in PALC. Most
interestingly, the reduced HETP measured in HILIC for caffeine never falls below 4. This suggests that the
mass transfer of caffeine between the multilayer adsorbed phase (due to the interactions of the strong
solvent and the silanol groups) and the acetonitrile-rich bulk eluent is slow.
© 2009 Elsevier B.V. All rights reserved.
1. Introduction
Hydrophilic interactions chromatography (HILIC) is an effective
alternative to reversed-phase liquid chromatography (RPLC) for the
separation of samples containing very polar compounds [1]. HILIC
is based on the distribution of the sample between a polar adsor-
bent surface (silica, diol, aminopropyl, and zwitterionic phases) and
a bulk aqueous phase, usually rich in organic solvent such as ace-
tonitrile [2,3]. While the mechanism of retention in HILIC is still
debated [4], the analysis of the elution times of water insoluble
samples [5] and the minor plateau perturbation method [6] showed
that a water-rich adsorbed multilayer (at least two layers) builds
up on the surface of polar stationary phases when acetonitrile-rich
(80–90%) bulk mobile phases are used. On the one hand, the more
polar the compound, the higher its concentration in the adsorbed
layers of water at equilibrium, suggesting that HILIC could have
a partition mechanism. On the other hand, the retention of polar
compounds increases with increasing water concentration in the
∗
Corresponding author. Tel.: +1 865 974 0733; fax: +1 865 974 2667.
E-mail address: guiochon@utk.edu (G. Guiochon).
mobile phase [7]. These observations demonstrate that partition
alone does not govern the retention mechanism of polar samples
in HILIC. Analytes compete with water for adsorption onto the polar
sites at the stationary phase surface (silanols of silica, amino and
zwiterrionic groups of silica-bonded phases).
The current shortage of acetonitrile forces scientists in academic
and industrial laboratories to consider replacing either acetonitrile
as a component of mobile phases or acetonitrile-consuming sep-
aration modes with other ones using different solvents. The first
option would be to use ethanol or propanol instead of acetoni-
trile. Ethanol has about the same elution strength as acetonitrile,
performs reasonably well, and costs much less, so it is a suitable
alternative solvent [8]. A second option is to take advantage of the
hydrophobic character of siloxane groups at the surface of silica
by using water-rich eluents in per aqueous liquid chromatogra-
phy (PALC) [7]. Although good retention factors can be achieved
in PALC, the adsorption mechanism is heterogeneous and involves
active adsorption sites, which results in column overloading tak-
ing place with rather small samples. In contrast, HILIC shows less
peak tailing upon overloading the column. This was illustrated
by the measurement of the adsorption isotherm of pyridine by
frontal analysis and by the calculation of its adsorption energy
0021-9673/$ – see front matter © 2009 Elsevier B.V. All rights reserved.
doi:10.1016/j.chroma.2009.12.004