Analytica Chimica Acta 537 (2005) 271–278
On-line emulsion formation and multi-element analysis of edible oils
by inductively coupled plasma atomic emission spectrometry
A.N. Anthemidis
∗
, V. Arvanitidis, J.A. Stratis
Laboratory of Analytical Chemistry, Department of Chemistry, Aristotle University, Thessaloniki 54124, Greece
Received 7 October 2004; received in revised form 4 January 2005; accepted 4 January 2005
Available online 15 February 2005
Abstract
A novel simple on-line emulsion formation system for multi-element analysis by inductively coupled plasma atomic emission spectrometry
(ICP-AES) was developed. A laboratory made magnetic-stirring micro-chamber was used for continuous edible emulsification with Triton
X-100 and for subsequent introduction of the emulsion into ICP-AES. The optimum concentration of oil in emulsion for maximum sensitivity
was 50% (v/v). The performance of the system was demonstrated for Ag, Al, B, Ba, Bi, Ca, Cd, Co, Cr, Cu, Fe, Ga, In, Mg, Mn, Ni, Pb, Tl and
Zn determination in olive, sunflower and corn oil. Good agreement was found between calibration curves for oil emulsion and aqueous ones
for the most of the investigated spectral lines. The calibration curves for each element were linear at least up to 1.0 mg l
-1
and the correlation
coefficients were ranged between 0.991 and 0.999. The proposed method was applied to the analysis of olive oil sample and compared with
ETAAS-wet digestion method. No significant differences were observed between the two methods.
© 2005 Elsevier B.V. All rights reserved.
Keywords: Inductively coupled plasma atomic emission spectrometry; On-line; Emulsion; Edible oil; Multi-element analysis
1. Introduction
The quality of edible oils with regard to freshness, stora-
bility and toxicity can be evaluated by the determination of
metals. Trace levels of metals like Fe, Cu, Ca, Mg, Co, Ni
and Mn are known to increase the rate of oil oxidation while
other elements such as Cr, Cd, and Pb are very important on
account of their toxicity and metabolic role. Thus, the de-
velopment of rapid and accurate analytical methods for trace
elements determination in edible oil has been a challenge in
quality control and food analysis.
Various techniques like electrothermal atomic absorp-
tion spectrometry (ETAAS) [1], inductively coupled plasma
atomic emission spectrometry (ICP-AES) [2] or mass spec-
trometry (ICPMS) [3,4] and voltammetry [5] have been em-
ployed for analysis of edible oils. However, sample pretreat-
ment procedures like: wet or dry digestion, microwave di-
∗
Corresponding author. Tel.: +30 2310 997707; fax: +30 2310 997719.
E-mail address: anthemid@chem.auth.gr (A.N. Anthemidis).
gestion, dilution with organic solvent and some extraction
methods are required, in order to eliminate the organic ma-
trix. These techniques are usually difficult to be automated,
time consuming, laborious and prone to contamination and
losses. On the other hand, direct analysis of samples with
high content of organic matrix by ICP techniques encoun-
ters many drawbacks and usually causes plasma to be ex-
tinguished. Another negative effect is the injector blocking
due to carbon deposition from incomplete oxidation of or-
ganic matrix [6]. An alternative technique for introduction of
oil samples directly into inductively coupled plasma (ICP) is
the on-line emulsification [2,7], which also offers some other
advantages like using aqueous standards solutions for cali-
bration and quantification, reduction of matrix interference,
higher sampling frequency and lower cost of analysis. Direct
introduction of oil samples in the form of emulsion into ICP
considerable simplifies the spray chamber and plasma torch
because no extra oxygen or complicated desolvation device is
needed. In this case, the use of stable emulsions with proper
surfactant concentration is very important. The on-line prepa-
0003-2670/$ – see front matter © 2005 Elsevier B.V. All rights reserved.
doi:10.1016/j.aca.2005.01.035