Characteristics and quality assurance of a dedicated open 0.23 T MRI
for radiation therapy simulation
Dennis Mah
a)
Department of Radiation Oncology, Fox Chase Cancer Center, Philadelphia, Pennsylvania19111
Michael Steckner
Philips Medical Systems, Cleveland, Ohio 44143
Elizabeth Palacio, Raj Mitra, Theresa Richardson, and Gerald E. Hanks
Department of Radiation Oncology, Fox Chase Cancer Center, Philadelphia, Pennsylvania19111
Received 27 December 2001; accepted for publication 8 August 2002; published 23 October 2002
A commercially available open MRI unit is under routine use for radiation therapy simulation. The
effects of a gradient distortion correction GDC program used to post process the images were
assessed by comparison with the known geometry of a phantom. The GDC reduced the magnitude
of the distortions at the periphery of the axial images from 12 mm to 2 mm horizontally along the
central axis and distortions exceeding 20 mm were reduced to as little as 2 mm at the image
periphery. Coronal and sagittal scans produced similar results. Coalescing these data into distortion
as a function of radial distance, we found that for radial distances of 10 cm, the distortion after
GDC was 2 mm and for radial distances up to 20 cm, the distortion was 5 mm. The dosimetric
errors resulting from homogeneous dose calculations with this level of distortion of the external
contour is 2%. A set of triangulation lasers has been added to establish a virtual isocenter for
convenient setup and marking of patients and phantoms. Repeated measurements of geometric
phantoms over several months showed variations in position between the virtual isocenter and the
magnetic isocenter were constrained to 2 mm. Additionally, the interscan variations of 12 ran-
domly selected points in space defined by a rectangular grid phantom was found to be within the
intraobserver error of 1 mm in the coronal, sagittal, and transverse planes. Thus, the open MRI
has sufficient geometric accuracy for most radiation therapy planning and is temporally stable.
© 2002 American Association of Physicists in Medicine. DOI: 10.1118/1.1513991
Key words: quality assurance, open MRI, treatment planning, MRI simulation, distortions
INTRODUCTION
For radiation treatments to be effective, the target volume
must receive the prescribed dose while keeping the dose to
surrounding tissues below their tolerance. Treatment plan-
ning and delivery consists of multiple steps
1
each contribut-
ing to the overall uncertainty. These uncertainties include
target definition, organ motion, setup error, accelerator out-
put, dose calculation, and biological response. Because of its
greater soft tissue contrast, MRI has been shown to provide
more consistent target delineation than CT for a variety of
sites
2–4
and MRI based planning has been shown to reduce
interphysician variation and produce a more accurate plan.
5
MRI based planning, usually via fusion of the MRI to CT is
used for treatment planning of cranial sites
2,6–9
and recently
has also been investigated
5
and routinely implemented for
the prostate.
10
However, the fusion process itself requires
additional time and effort and may add uncertainty from the
imperfect registration of the two image sets.
11
CT and MRI
provide largely redundant anatomical information, but due to
concerns about MRI
12
image distortion, both scans are nor-
mally acquired for treatment planning. We and other investi-
gators are attempting to determine if just an MRI scan is
sufficient for treatment planning, i.e., if an MRI simulator
can be developed.
12,13
For MRI to replace CT there are a variety of challenges
that must be overcome. Some of these were initially de-
scribed by Fraass et al.,
12
and include production of digitally
reconstructed radiographs DRRs, development of appropri-
ate MRI treatment planning software with inhomogeneity
corrections, and correction or at least quantification of dis-
tortions due to chemical shift, susceptibility, and field gradi-
ents. Many of these challenges are the subject of active re-
search. For instance, Ramsey et al.
14
have considered the
possibility of producing DRRs using a T1 weighted pulse
sequence. Another approach is to contour the bones on the
cross sectional slices to produce a DRR.
15
Recently, Frans-
son et al.
16
have suggested using bulk inhomogeneities e.g.,
setting lung to a density of 0.3 g/cc for dose calculation in
MRI data sets.
Distortions in MRI arise from both the MRI and from
patient induced effects. Patient induced effects include sus-
ceptibility and chemical shift distortions, which for low field
i.e., 0.2 T magnets are limited to 1 pixel,
17
implying
that the distortions of internal anatomy for these MRIs are
clinically acceptable. Distortions arising from the magnet it-
self fall into two categories: field inhomogeneity and nonlin-
ear gradients. Since MRI images derive spatial information
from the spatially varying frequencies of the signals, any
distortion in the main field ( B
0
) maps into spatial uncer-
2541 2541 Med. Phys. 29 „11…, November 2002 0094-2405Õ2002Õ29„11…Õ2541Õ7Õ$19.00 © 2002 Am. Assoc. Phys. Med.