Analysis of Ultrahigh Molecular Weight Polyethylene Failure in
Artificial Knee Joints: Thermal Effect on Long-Term Performance
Tai-Horng Young,
1
Cheng-Kung Cheng,
2
Ye-Ming Lee,
3
Li-Yen Chen,
1
Chun-Hsiung Huang
3
1
Graduate Institute of Biomedical Engineering, College of Medicine and College of Engineering, National Taiwan University,
Taipei, Taiwan, Republic of China
2
Institute of Biomedical Engineering, National Yang Ming University, Taipei, Taiwan, Republic of China
3
Department of Orthopaedic Surgery, Mackay Memorial Hospital, Taipei, Taiwan, Republic of China
Received 6 October 1997; accepted 24 April 1998
Abstract: The mechanism resulting in damage to and failure of ultrahigh molecular weight
polyethylene (UHMWPE) tibial inserts was investigated on clinically retrieved components.
The severity of the subsurface damage increased with the length of time that the component
had been implanted. A theoretical analysis was developed to account for the generation of
subsurface damage based on a heat transfer model. Friction generates surface heat during
articulation of total knee systems. Due to the cooling effect of body fluid on the surface, the rise
in temperature on the UHMWPE surface is lower than that below the surface. The peak
temperature was estimated to occur on a plane positioned about 1 to 2 mm below the surface.
This result was similar to the bulk temperature variation observed during in vivo and in vitro
studies by other investigators. Although the difference in temperature on and below the
surface is only a few degrees, the thermal effect becomes apparent after a long time and may
be explained by the viscoelastic behavior of polymers: the temperature–time equivalence. It is
therefore suggested that this thermal effect is another contributory factor to material damage,
in addition to high stress and oxidative degradation (in appropriate cases). Therefore, any
technological efforts aimed at improving the performance of artificial joint prostheses should
minimize the thermal effects at the subsurface of the articular components. © 1999 John Wiley
& Sons, Inc. J Biomed Mater Res (Appl Biomater) 48: 159 –164, 1999
Keywords: UHMWPE; subsurface yellowing; thermal effect
INTRODUCTION
The long-term performance of the articulating system and
materials in the design of artificial joints is very important,
particularly for younger, heavier, and more active patients.
One of the most significant factors that leads to revision in
total joint arthroplasty is the damage to and failure of ultra-
high molecular weight polyethylene (UHMWPE).
1–4
Although many factors can affect the performance of
UHMWPE in artificial joints, most investigators have
adopted a purely mechanical force approach (e.g., estimation
of stress distribution or trials to increase the strength of
materials).
5–10
In addition, postirradiation oxidative degrada-
tion of UHMWPE is also receiving attention as a potential
factor influencing the performance of UHMWPE.
11–14
However, one aspect previously ignored with regard to long-
term performance is the tendency for the articulating systems
to generate frictional heat. A recent study performed by Tepic
et al. showed that there was a tendency for the temperature in
the femoral head and acetabular cartilage to increase by about
2.5°C in intact cadaver hip joints.
15
Davidson et al. showed
that heat generated at the articulating surface increases the
temperature of UHMWPE by 5 to 10°C in a simulator.
16 –18
In vivo measurements with a telemeterized hip implant by
Rohlmann et al. showed an increase in the temperature of
3.8°C in the prosthetic neck of the hip joint when the patient
walked outside for 90 min.
19
Because frictional heating can
increase the creep, wear, and oxidation degradation of
UHMWPE, the associated property changes and performance
at elevated temperatures may be an important factor to be
considered. However, a complete explanation of the effect of
the frictional heating on UHMWPE and its implications seem
to be lacking. Therefore, the objectives of this study were to
summarize the experimental observations from retrieved
UHMWPE tibial inserts and to present a theoretical explana-
tion for these results.
Correspondence to: Dr. Chun-Hsiung Huang (e-mail: thyoung@ntumc1.mc.ntu.
edu.tw)
© 1999 John Wiley & Sons, Inc. CCC 0021-9304/99/020159-06
159