Sunita P. Ho
Department of Preventive and Restorative Dental
Sciences,
University of California San Francisco
San Francisco, CA 94143
Paul F. Joseph*
Department of Mechanical Engineering
Clemson University
Clemson, SC 29634
Michael J. Drews
School of Materials Science and Engineering
Clemson University
Clemson, SC 29634
Thomas Boland
Martine LaBerge
Department of Bioengineering
Clemson University
Clemson, SC 29634
Experimental and Numerical
Modeling of Variable Friction
Between Nanoregions in
Conventional and Crosslinked
UHMWPE
Recently, highly crosslinked UHMWPE components have been promoted for their high
abrasive wear resistance over conventional UHMWPE (PE) in total joint replacement
(TJR) prostheses to minimize osteolysis and consequent implant loosening. This study was
aimed at investigating the role of friction gradients induced by localized coefficients of
friction at both crystalline and amorphous nanoregions in PE, and crystalline and
crosslinked nanoregions in crosslinked UHMWPE (XPE), in submicron wear debris gen-
eration. An abrasive wear study performed on both XPE and PE using atomic force
microscopy (AFM) illustrated that the onset of plastic deformation for XPE occurred at a
normal load that was approximately 3 times higher when compared to PE. Coefficients of
friction
d
of 0.2, 0.35, and 0.61, experimentally derived using AFM, were used as
representative
d
for crystalline, amorphous, and crosslinked nanoregions, respectively,
in a numerical Hertzian model. An increase in (0.20.02, 0.350.01 and 0.60.04)
was observed with a decrease in crystallinity and storage modulus at 22°C. Using the
Hertzian contact model, it was observed that variability in friction between nanoregions
contributed to higher magnitude stresses for XPE (0.2 to 0.61; maximum
eff
=2.8 com-
pared to PE (0.2 to 0.35; maximum
eff
=1.1 over a negligible thickness of the interfa-
cial zone (IZ) between nanoregions. The experimentally observed increase in abrasive
wear resistance of XPE could be attributed to an increase in the thickness of the interfa-
cial zone between nanoregions with changing gradually from crystalline to crosslinked
nanoregions, a situation that may not be observed with PE. This would cause a decrease
in the friction gradient and resulting stresses thereby agreeing with the observed experi-
mental higher abrasive wear resistance for XPE. However, in both PE and XPE, the
presence of stress concentrations over a period of time could lead to irreversible damage
of the material eventually generating submicron wear debris. Hence, semicrystalline,
inhomogenous UHMWPE with several nanoregions (amorphous and crystalline) would
be at a disadvantage for bearing application in terms of abrasive wear resistance com-
pared to UHMWPE with relatively lower number of nanoregions and crosslinked
nanoregions. DOI: 10.1115/1.1645530
Introduction
Total joint replacement TJR prostheses consist of metallic and
nonmetallic components such as mirror-finish cast cobalt-
chromium Co-Cr alloy or ceramic articulating on a metallic,
ceramic or a polymeric ultra high molecular weight polyethylene
UHMWPE counter components 1,2. UHMWPE has been the
principal polymeric material used in artificial joints since the early
1960’s to replace damaged articular cartilage in injured and dis-
eased synovial joints 1,2. Despite the many advantages and wide
use of UHMWPE, the generation of UHMWPE submicron or
nanoscale wear debris, in TJR prosthesis has been of a primary
concern for the past four decades 3,4. The submicron UHMWPE
wear particles have been observed clinically to accumulate in the
periprosthetic tissues. This accumulation causes an inflammatory
reaction, leading to a cellular release of cytokines inducing bone
resorption, and consequently loss of fixation of the device and
ultimately device failure 5,6.
The combined effort of in-vitro joint simulation and computa-
tional modeling studies has contributed to elucidating the cause of
different wear mechanisms in UHMWPE at a macroscopic scale
7,8. Wang and collaborators suggested that large scale deforma-
tions, texture evolution, fracture and surface rupture within the
surface region of UHMWPE components over repeated cyclic
motion could cause macroscopic failure in the material eventually
leading to total failure of the implant 9. Although these studies
provided an insight to the wear processes in TJR prostheses, they
do not explain the initiating cause of the submicron wear debris
generation, a nanoscale phenomenon. Additionally, in the numeri-
cal models, the coefficient of friction at the tribological contact,
was either eliminated or assumed a constant.
In the studies focused on identifying the initiating cause of
failure, there is now a transition occurring in the field of materials
analysis from continuum macro scale to discrete asperity-to-
asperity nanoscale contact 10 which could provide new insight
to the basic problem of submicron UHMWPE submicron wear
debris. It is well documented to date that when two engineering
surfaces come into contact, plastic deformation may occur at the
*Corresponding author: Department of Mechanical Engineering, Clemson Uni-
versity, Box 340921, Clemson, SC 29634-021. Office: 235 Fluor Daniel Engineering
Innovation Building. Phone: 864 656-0545, Fax: 864 656-4435, e-mail:
paul.joseph@ces.clemson.edu
Contributed by the Bioengineering Division for publication in the JOURNAL OF
BIOMECHANICAL ENGINEERING. Manuscript received by the Bioengineering Divi-
sion February 25, 2003; revision received October 14, 2003. Associate Editor:
M. Toner.
Copyright © 2004 by ASME Journal of Biomechanical Engineering FEBRUARY 2004, Vol. 126 Õ 111