0885–3010/$25.00
©
2011 IEEE
316 IEEE TRANSACTIONS ON ULTRASONICS, FERROELECTRICS, AND FREQUENCY CONTROL, . 58, . 2, FEBRUARY 2011
Abstract—The frequency response of ultrasonic detectors
is commonly calibrated by finding their sensitivity to incident
plane waves at discrete frequencies. For certain applications,
such as the emerging field of optoacoustic tomography, it is
the response to point sources emitting broadband spectra that
needs to be found instead. Although these two distinct sensi-
tivity characteristics are interchangeable in the case of a flat
detector and a point source at infinity, it is not the case for
detectors with size considerably larger than the acoustic wave-
length of interest or those having a focused aperture. Such
geometries, which are common in optoacoustics, require direct
calibration of the acoustic detector using a point source placed
in the relevant position. In this paper, we report on novel
cross-validating optoacoustic methods for measuring the fre-
quency response of wideband acoustic sensors. The approach
developed does not require pre-calibrated hydrophones and
therefore can be readily adopted in any existing optoacoustic
measurement configuration. The methods are successfully con-
firmed experimentally by measuring the frequency response of
a common piezoelectric detector having a cylindrically focused
shape.
I. I
M
popular piezoelectric ultrasonic detectors at-
tain relatively limited bandwidth and non-uniform
frequency characteristics, leading to distortion of the de-
tected acoustic signals. In imaging applications, signal
distortions are translated into image artifacts and loss of
accuracy. Such artifacts can be reduced if the detector
characteristics which are relevant to the imaging modal-
ity are known and corrected for. Thus, the calibration of
ultrasonic detectors is not only important for assessing
image fidelity, but may also be used to improve the overall
quantification abilities.
Several calibration methods have been developed using
acoustic plane waves [1]–[6]. One of the early methods
employed was the reciprocity method, in which the same
transducer or transducers is used both as the source and
as the detector [1]–[3]. Using this technique, the efficiency
of the transducer in both modes is embodied in the mea-
sured acoustic signal. The known relation between these
two efficiencies, manifested in the reciprocity parameter, is
then utilized to extract both efficiencies from the measure-
ment. One of the advantages of the reciprocity method is
that it does not require using a primary standard, i.e., an
acoustic detector whose response is known. Another ap-
proach which avoids using a primary standard is to mea-
sure the driving voltage fed into a transducer and use a
modeling approach to calculate the formed acoustic field
[4]. The calibration is performed by placing a hydrophone
in front of the transducer and comparing its measured
signal to the one predicted by the model. When a primary
standard is used, the calibration of acoustic detectors is
performed by creating a wide-band acoustic field and mea-
suring it with both the primary standard and the sensor
that is to be calibrated. By comparing the two detected
signals, the frequency response of the detector can be cal-
culated. The acoustic fields can be created by wide-band
transducers [5], [6], or by nonlinear acoustic propagation
effects [7]. The primary standard may be a pre-calibrated
hydrophone [7] or a setup based on optical interferometry
[6], [7].
In recent years, there has been a growing interest in
new imaging modalities based on the thermoacoustic ef-
fect, which require calibrated acoustic sensors to perform
quantified imaging [8]–[14]. In optoacoustic tomography,
for instance, ultra-wideband acoustic fields are generated
by thermal expansion of tissue exposed to high-power short
laser pulses. Here, the imaged object acts as an acoustic
(or optoacoustic) source. Because the detection is acous-
tic, optoacoustic tomography is capable of mapping opti-
cal contrast while attaining diffraction-limited ultrasonic
resolution, unaffected by light diffusion. Furthermore,
techniques like multispectral optoacoustic tomography
(MSOT), in which the laser operates at multiple wave-
lengths, extend this mapping capability to imaging and
mapping bio-distribution of spectrally-distinct molecular
biomarkers, also attracting a great deal of interest from
the biological and medical communities [13], [14]. Optoa-
coustic imaging typically operates with much wider-band
signals compared with conventional ultrasound imaging;
therefore, the detector characterization needs to be per-
formed over wider frequency bands to attain accurate im-
age quantification.
One of the common detection geometries in optoacous-
tic tomography is based on transducers cylindrically fo-
cused in the detection plane [8]–[12]. Because of their large
detection area, the use of focused sensors significantly im-
Optoacoustic Methods for Frequency
Calibration of Ultrasonic Sensors
Amir Rosenthal, Vasilis Ntziachristos, and Daniel Razansky
Manuscript received July 15, 2010; accepted November 4, 2010. A.
R. acknowledges the financial support of the European Community’s
Seventh Framework Programme (FP7/2007-2013 under grant agreement
number 235689). D. R. acknowledges support from the German Research
Foundation (DFG) Research Grant (RA 1848/1) and the European Re-
search Council Starting Grant. V. N. acknowledges financial support
from the European Research Council Advanced Investigator Award, and
the BMBF’s Innovation in Medicine Award.
The authors are with the Institute for Biological and Medical Imaging
(IBMI), Technical University of Munich and Helmholtz Center Munich,
Neuherberg, Germany (e-mail: eeamir@gmail.com).
A. Rosenthal is also with the Cardiovascular Research Center (CVRC)
and Cardiology Division, Massachusetts General Hospital and Harvard
Medical School, Boston MA.
Digital Object Identifier 10.1109/TUFFC.2011.1809