Please cite this article in press as: Raamat R, et al. Comparison of oscillometric pulse amplitude envelopes recorded from the locally compressed
radial arteries. Med Eng Phys (2010), doi:10.1016/j.medengphy.2010.08.002
ARTICLE IN PRESS
G Model
JJBE-1731; No. of Pages 7
Medical Engineering & Physics xxx (2010) xxx–xxx
Contents lists available at ScienceDirect
Medical Engineering & Physics
journal homepage: www.elsevier.com/locate/medengphy
Comparison of oscillometric pulse amplitude envelopes recorded from the
locally compressed radial arteries
R. Raamat
∗
, J. Talts, K. Jagomägi, J. Kivastik
Department of Physiology, University of Tartu, Tartu, Estonia
article info
Article history:
Received 7 March 2010
Received in revised form 1 August 2010
Accepted 7 August 2010
Keywords:
Oscillometric envelope
Oscillometric pattern
Oscillogram
Radial non-invasive blood pressure
Photoplethysmographic
Pneumoplethysmographic
abstract
Simultaneously recorded oscillometric envelopes, obtained pneumo- and photoplethysmographically
from a small local pad-type pneumatic cuff on the left and from a larger cuff on the right radial artery, were
compared in 17 healthy subjects. For oscillometric estimation, specific fixed ratios based on evidence in
the literature were used. The obtained envelopes for each person were shifted and aligned at the point of
upper arm mean arterial pressure for this person, thus eliminating the brachial-to-radial mean blood pres-
sure gradient and possible left–right difference. In this way, the shape of differently recorded envelopes
as a determinant of the accuracy of oscillometric estimation was studied. Results showed an advan-
tage of photoplethysmographically compared to pneumoplethysmographically recorded envelopes. For
a smaller cuff (diameter 40 mm), the mean difference in mm Hg ‘oscillometric estimate minus auscul-
tatory reference’ and standard deviation were in the case of photo recording for systolic and diastolic
pressures -0.6 (6.3) and 1.2 (3.4), respectively. In the case of pneumo recording, these parameters were
considerably larger, being 12.1 (11.9) and -6.2 (10.9), respectively. For a larger cuff the same tendency was
revealed. Photo recording was found to be less sensitive to alterations in the cuff size and characteristic
ratios.
© 2010 IPEM. Published by Elsevier Ltd. All rights reserved.
1. Introduction
The oscillometric method is increasingly used in blood pres-
sure (BP) measurement. Due to the simplicity and reliability of this
method, the oscillometric principle is employed in the majority
of present-day automatic and semiautomatic non-invasive blood
pressure monitors.
Oscillometric blood pressures are typically determined from
the envelope of successive oscillometric pulse amplitudes obtained
from the occlusive cuff during its inflation or deflation. The highest
point of the envelope curve is generally regarded as the mean arte-
rial pressure (MAP) [1,2]. This is consistent with the oscillometric
maximum criterion described by Marey more than 130 years ago
[3].
Two general types of criteria have been used to estimate systolic
(SBP) and diastolic (DBP) blood pressures: height-based criteria and
slope-based criteria [4]. In the height-based approach the systolic
and diastolic pressures are determined using special fractions of the
maximum oscillation amplitude [5,6]. These fractions are known
as characteristic ratios (systolic and diastolic, respectively). In the
slope-based approach the cuff pressure at which the oscillometric
pulse amplitude increases rapidly is taken as the systolic pressure,
∗
Corresponding author. Tel.: +372 737 4987.
E-mail address: raamat@ut.ee (R. Raamat).
while that at which the amplitude decreases rapidly is taken as
the diastolic pressure. Mathematically, the latter criterion defines
the points at which the first derivative of the envelope curve is
maximum or minimum.
There are no definitive correct values for the systolic and dias-
tolic characteristic ratios, but rather that different manufactures
use different values. Theoretical and experimental work has sug-
gested the following values for characteristic ratios:
(a) systolic from 0.46 to 0.64 and diastolic from 0.59 to 0.80 (theo-
retical work by Ursino and Cristalli [7]),
(b) systolic 0.593 and diastolic 0.717 (theoretical work by
Drzewiecki et al. [6]),
(c) systolic from 0.45 to 0.57 and diastolic from 0.69 to 0.89 (exper-
imental work by Geddes et al. [5]),
(d) systolic (mean and SD) equal to 0.49 (0.11) and diastolic equal
to 0.72 (0.12) (experimental work by Amoore et al. [8]).
Specific studies have demonstrated an effect of the shapes
of oscillometric pulse amplitude envelopes on the differences
between auscultatory and oscillometric BP measurement [8,9].
Subsequently, recording of oscillometric envelopes of appropriate
shape and stability remains a topical issue for the oscillometric
method.
Oscillometric wrist monitors have shown lower reliability com-
pared to upper arm devices. On the other hand, oscillometric wrist
1350-4533/$ – see front matter © 2010 IPEM. Published by Elsevier Ltd. All rights reserved.
doi:10.1016/j.medengphy.2010.08.002