Talanta 80 (2010) 1445–1451
Contents lists available at ScienceDirect
Talanta
journal homepage: www.elsevier.com/locate/talanta
Analysis of commercial beverage products by size exclusion
chromatography coupled with UV–vis absorbance detection
and dynamic surface tension detection
Karisa M. Pierce
a,∗
, Emilia Bramanti
b
, Massimo Onor
b
, Roberto Spiniello
b
, Alexandra Kangas
a
,
Kristen J. Skogerboe
c
, Robert E. Synovec
d
a
Department of Chemistry and Biochemistry, 3307 Third Avenue West, Suite 205, Seattle Pacific University, Seattle, WA 98119, USA
b
National Research Council of Italy (CNR), Istituto per i Processi Chimico-Fisici, Laboratory of Instrumental Analytical Chemistry, Via G. Moruzzi 1, 56124 Pisa, Italy
c
Department of Chemistry, Seattle University, Seattle, WA 98122, USA
d
Department of Chemistry, Box 351700, Center for Process Analytical Chemistry, University of Washington, Seattle, WA 98195, USA
article info
Article history:
Received 16 June 2009
Received in revised form
24 September 2009
Accepted 25 September 2009
Available online 3 October 2009
Keywords:
Instant coffee
Surface tension
Tetrabutylammonium (TBA)
Liquid chromatography
Proteins
abstract
Multidimensional analysis of instant coffee and barley beverage samples using size exclusion chro-
matography (SEC) combined with a dynamic surface tension detector (DSTD) and a UV–vis absorbance
detector (UV) is reported. A unique finding of this study was the action of the tetrabutylammonium
(TBA) cation as a modifying agent (with bromide as the counter anion) that substantially increased
the surface pressure signal and sensitivity of many of the proteins in the chromatographically sepa-
rated samples. The tetrabutylammonium bromide (TBAB) enhancement of the surface pressure signal
was further investigated by studying the response of 12 commercial standard proteins (-lactalbumin,
-lactoglobulin, human serum albumin (HSA), albumin from chicken egg white (OVA), bovine serum
albumin (BSA), hemoglobin, -chymotrypsinogen A, cytochrome C, myoglobin, RNase A, carbonic
anhydrase, and lysozyme) in buffer performed using flow injection analysis (FIA) coupled with the
DSTD with and without various concentrations of TBAB. The FIA-DSTD data show that 1 mM TBAB
enhances sensitivity of HSA detection, by lowering the limit of detection (LOD) from 2 mg/mL to
0.1 mg/mL. Similarly, the LOD for BSA was reduced from 1 mg/mL to 0.2 mg/mL. These FIA-DSTD
experiments allowed the detection conditions to be optimized for further SEC-UV/DSTD experiments.
Thus, the SEC-UV/DSTD system has been optimized and successfully applied to the selective analy-
sis of surface-active protein fractions in a commercial instant coffee sample and in a soluble barley
sample. The complementary selectivity of using the DSTD relative to an absorbance detector is also
demonstrated.
© 2009 Elsevier B.V. All rights reserved.
1. Introduction
The direct measurement of surface tension is of interest in the
investigation of molecular interactions at the air–liquid interface,
the study of thermodynamic kinetics of protein denaturation at
the air–liquid interface, and in determining foaming capabilities of
foods and beverages. Proteins play a key role in the stabilization of
foams, emulsions, and composite systems in food products and in
cosmetics [1–7]. The creation of emulsions and foams in food prod-
ucts represent dynamic processes and the dynamic surface-active
properties of protein solutions are considered to play an important
role in the formation and stabilization of these systems [6,8,9].
∗
Corresponding author. Tel.: +1 206 281 2102.
E-mail address: pierck1@spu.edu (K.M. Pierce).
Traditional methods of quantifying surface tension are based
on optical observations of drop growth, optical imaging, ten-
siometry, and manual observations of foam performance [10–12].
Most of these techniques are restricted to static measurements
and cannot assess the dynamic nature of foams and emulsions.
Additionally there may be other limitations of these methods
such as protein adsorption [13] or lack of compatibility with
flow injection analysis (FIA) or liquid chromatographic systems
[14].
The limitations of previous methods have led to the devel-
opment of a versatile technique to assess the dynamic surface
pressure of flowing liquids. The dynamic surface tension detector
(DSTD) is a high-throughput automated chemical analyzer capable
of quantifying the surface tension of a complex sample [13–17].
The surface-tension-lowering response observed by the DSTD is
converted to a quantity known as the surface pressure. The DSTD
provides real-time surface pressure data for surface-active sample
0039-9140/$ – see front matter © 2009 Elsevier B.V. All rights reserved.
doi:10.1016/j.talanta.2009.09.051