An electrothermal vaporization unit with axially focusing convection
upstream and influence of modifiers. Part I: Experimental
☆
A. Trenin
a,b
, M. Gafurov
a,b
, A.Kh. Gilmutdinov
b
, G. Hermann
a,
⁎
a
I. Physikalisches Institut, Justus-Liebig-Universität Giessen, Heinrich-Buff-Ring 16, 35392 Giessen, Germany
b
Kazan State University, Kremlevskaja Str. 18, 420008 Kazan, Russia
Received 19 September 2006; accepted 3 November 2006
Available online 18 December 2006
Abstract
A novel, further improved construction of an electrothermal vaporization (ETV) unit is described, and new measurements are presented. Its
main feature is an upstream within a vertical tube creating an axially focusing convection zone, where the condensation occurs in a laminar flow
predominantly apart from the colder walls. The upstream is shielded against incandescent emission of the graphite furnace (GF) and surrounded
with different cooling gas flows, that have additional focusing effect. The transported ETV generated aerosol is electrostatically precipitated onto a
secondary platform that is subsequently analyzed for the transported analytes. Transport efficiencies (TEs) for Cu, Fe, Mn, Ni, and Pb are studied
using a continuum source coherent forward scattering multi-element spectrometer coupled to the ETV. Increasing TE is observed with the ageing
of the GF. The effect is presumably ascribed to the growing content of carbon particulates released during the vaporization. A positive impact of K
and Pd modifiers in microgram amounts foremost on the TEs of more volatile species is observed and discussed.
© 2006 Elsevier B.V. All rights reserved.
Keywords: Solid sampling; Electrothermal vaporization; Electrostatic precipitation; Transport efficiency; Simultaneous multielement determination
1. Introduction
Various constructions of electrothermal vaporization (ETV)
cells have been used as sample introduction tools for
inductively coupled plasma optical emission and mass spec-
trometry (ICP-OES/MS) [1–5] as well as for graphite furnace
(GF) analytical techniques [6–9]. Advantages of the ETV
dosing are the capability for solid, liquid, and slurry sampling,
reduction of sample preparation time with reduced risk of
sample contamination and of analyte loss prior to analysis. In
addition, the ETV offers the possibility to separate the analytes
from matrix components by external thermal pretreatment and
vaporization steps.
Up to now, the ETV devices are based on the following designs:
(1) cup vaporizers, (2) end-on flow-through GFs with vapor outlet
through one of the ends of the graphite tube [2–8,10–12], and
(3) upstream configurations with the gas entering the GF through
its ends and released upwards through a hole in the tube center
[1,9,12,13]. The comparison shows the superiority of the upstream
configuration [12]. Nevertheless, for both constructions an evident
dependence of analyte transport efficiency (TE) and analyte losses
on element volatility, sample mass, ETV operation conditions and
other parameters is revealed.
The application of modifiers such as NaCl [10], Mg/Pd(NO
3
)
2
[14,15], citric acid [14], or components of diluted NASS-3 sea-
water [16] can support the formation of transportable aerosol and
increase the TE that allows to reduce the above mentioned
dependencies. Generally, the modifier may improve the formation
of carriers and the co-vaporization of sample analytes with
carriers. Gaseous additives such as carbon hydrates, halide carbon
hydrates [5,12,17,18], and carbon particle aerosols [19] are used
Spectrochimica Acta Part B 62 (2007) 231 – 241
www.elsevier.com/locate/sab
☆
This paper was presented at the VII European Furnace Symposium on
Atomic Absorption Spectrometry, Electrothermal Vaporization and Atomization
(VII EFS) and XII Solid Sampling Colloquium with Atomic Spectrometry (XII
SSC), held, in St. Petersburg, Russia, July 2–7, 2006 and is published in the
special issue of Spectrochimica Acta Part B, dedicated to that conference.
⁎