Real Time Online Correction of Mass Shifts and Intensity Fluctuations
in Extractive Electrospray Ionization Mass Spectrometry
Yong Tian,
†
Miao Yu,
†
Jian Chen,
‡
Chunxiao Liu,
†
Jianbo Shi,*
,†
Huanwen Chen,*
,‡
and Guibin Jiang
†
†
State Key Laboratory of Environmental Chemistry and Ecotoxicology, Research Center for Eco-Environmental Sciences, Chinese
Academy of Sciences, Beijing 100085, China
‡
Jiangxi Key Laboratory for Mass Spectrometry and Instrumentation, College of Chemistry, Biology and Material Sciences, East
China University of Technology, Nanchang, Jiangxi 330013, China
* S Supporting Information
ABSTRACT: Real time online calibration of mass shift and intensity
fluctuation to improve the accuracy of measurements for identification
and quantitation in trace mass spectrometric analysis was demon-
strated using extractive electrospray ionization mass spectrometry
(EESI-MS). The signals of authentic compounds (e.g., lysine (Lys),
proline (Pro), and histidine (His)) spiked into the extractive solution
for the EESI process were used as the references to calibrate the signal
of analytes (e.g., methionine (Met)) in the untreated sample solution.
The intensity of the analyte signal was recorded simultaneously with
the reference signals. The analyte signals at a given time point were
calibrated on the basis of these correlation factors and real time signal
response of the reference. The calibrated signal of Met at 10 μgL
-1
was improved with a better signal-to-noise ratio (S/N from 2.3 to 4.3),
better linearity (R
2
from 0.9758 to 0.9980), and reduced relative
standard deviation (RSD from 9.8% to 6.0%). The shift of mass-to-charge ratio of Met signal between the detected and
theoretical values was decreased from 247 ± 133 to -7 ± 167 ppm for 50 min of detection using a linear ion trap mass analyzer
and was reduced from -0.27 ± 0.60 to -0.12 ± 0.23 ppm for 50 min of detection using an Orbitrap mass analyzer (P = 95%).
This method has been validated using a certified standard amino acids solution (GBW(E)100062) and applied for quantitative
detection of amino acids in chicken feed, urine, nutritional drink, and facial mask samples, showing that the method is useful to
improve the accuracy of mass spectrometric analysis.
M
ass spectrometry (MS) is increasingly used for trace
detection in multiple disciplines such as element
analysis,
1
clinical chemistry,
2
metallomics,
3
proteomics,
4
metabolomics,
5
lipidomics,
6
etc. due to its unparalleled
capability for identification and quantification of the analytes.
Normally, efforts have been spent to improve the sensitivity
and resolution power of MS instruments to acquire more
accurate and reliable results.
7-9
To date, most commercial mass
spectrometers are routinely calibrated prior to use for sample
analysis to guarantee the accuracy and precision of the
measurement,
10
especially for cases (e.g., analysis of biological,
geological, and environmental samples) where matrix effects are
notably detected. Because the spectral baseline shift is normally
encountered,
11,12
the long-term accuracy and reliability of the
analytical results can be affected after a long operation time.
This problem is particularly serious for biological analyses using
large amount of samples, e.g., metabolomics,
11,13
causing
misinterpretation of experimental data.
External calibration, standard addition, and internal standards
are commonly used to obtain quantitative results.
12
However,
the instrument drift is hardly corrected using these methods in
practical applications.
14
External calibration performed with MS
instruments with high frequency should be able to reduce the
instrument shift, but it is too laborious to carry out, especially in
cases where online real time monitoring or a large batch of
sample analysis is required.
10
Isotope labeled internal standards
are also widely used to correct the signal variation caused by
differences in sample composition and instrument drift.
15,16
Unfortunately, the isotope labeled compounds are usually
expensive and not commonly available for all analytes. For
certain compounds, the naturally occurring isotopes of the
analyte and nonlinear signal response would also influence the
signal of the internal standard.
11,17,18
Eckers et al.
19
and Wolff
et al.
20
developed the LockSpray for accurate mass measure-
ments. By switching the reference and sample spray alternately
into MS, the mass value of the analytes could be calibrated by
the reference for both MS and tandem mass spectrometry
(MS/MS) measurements.
19-22
However, the mass intensity
correction by two sprayers was not reported to our knowledge.
Therefore, it is highly desirable to develop a new method for
Received: August 25, 2015
Accepted: November 23, 2015
Published: November 23, 2015
Technical Note
pubs.acs.org/ac
© 2015 American Chemical Society 11962 DOI: 10.1021/acs.analchem.5b04372
Anal. Chem. 2015, 87, 11962-11966