ANALYST
FULL PAPER
THE
www.rsc.org/analyst
Micro-machined planar field asymmetric ion mobility
spectrometer as a gas chromatographic detector
G. A. Eiceman,*
a
E. G. Nazarov,
a
R. A. Miller,
b
E. V. Krylov
a
and A. M. Zapata
c
a
Department of Chemistry and Biochemistry, New Mexico State University, Las Cruces, NM
88003, USA
b
Sionex Corporation, Wellesley Hills, MA 02481, USA
c
Draper Laboratory, Cambridge, MA, USA
Received 18th December 2001, Accepted 1st March 2002
First published as an Advance Article on the web 20th March 2002
A planar high field asymmetric waveform ion mobility spectrometer (PFAIMS) with a micro-machined drift tube
was characterized as a detector for capillary gas chromatography. The performance of the PFAIMS was compared
directly to that of a flame ionization detector (FID) for the separation of a ketone mixture from butanone to
decanone. Effluent from the column was continuously sampled by the detector and mobility scans could be
obtained throughout the chromatographic analysis providing chemical information in mobility scans orthogonal to
retention time. Limits of detection were approximately 1 ng for measurement of positive ions and were
comparable or slightly better than those for the FID. Direct comparison of calibration curves for the FAIMS and
the FID was possible over four orders of magnitude with a semi-log plot. The concentration dependence of the
PFAIMS mobility scans showed the dependence between ion intensity and ion clustering, evident in other
mobility spectrometers and atmospheric pressure ionization technologies. Ions were identified using mass
spectrometry as the protonated monomer and the proton bound dimer of the ketones. Residence time for column
effluent in the PFAIMS was calculated as ~ 1 ms and a 36% increase in extra-column broadening versus the FID
occurred with the PFAIMS.
Introduction
Detectors in gas chromatographic separations have long
supplemented the chromatographic performance of columns by
providing additional information related to a sample. These may
be either enhanced selectivity of sample ionization (electron
capture detector, surface ionization detector) or additional
chemical–structural information about a component of a sample
(mass selective detector).
1–3
Though a remarkably high level of
column efficiency can be obtained with commercially available
bonded-phase capillary columns, the widespread utilization of
mass spectrometers as a GC detector in environmental,
industrial, and medical measurements is evidence that addi-
tional dimensions of information in GC separations have
importance. Thus, plots of ion abundance versus m/z orthogonal
to retention time often add analytical confidence sufficient to
justify the high costs of purchase, operation, and maintenance of
mass spectrometers as GC detectors.
In 1982, a conventional IMS drift tube design was described
for use as a GC detector for capillary columns, where sample
clearance in the source region was fast and sample neutrals were
prevented from diffusing into the drift region.
4
This detector
exhibited high-speed response, low memory effects and repro-
ducible gas phase ion chemistry inside the drift tube. Recently,
class specific information has been discovered in mobility
spectra under certain conditions of low moisture and high
temperature.
5,6
Consequently, mobility spectrometers can be
adjusted to provide spectra either with intact product ions or
with fragment ions so that information density might be
controlled by the analyst. In such conditions, confidence levels
in categorizing mobility spectra by chemical class may be very
high (for chemicals class 90–95%, for chemical identification
inside classes ~ 80%). In summary, the information density in
mobility spectra is sufficiently high so that mobility spectrome-
ters may be considered economical and sensible alternatives to
a mass spectrometer as GC detectors when utilities, size, weight
and cost are restrictions. One example is air quality monitoring
on-board the international space station where measurements
are now made using a GC/IMS instrument, the Volatile Organic
Analyzer.
7
A significant limitation to the widespread use of the
conventional mobility spectrometers as GC detectors has been
the cost of manufacturing drift tubes. Compared to mass
spectrometers, mobility spectrometers are simple and inex-
pensive; however, traditional mobility detectors can be con-
sidered costly compared with the FID or a thermal conductivity
detector. Though miniaturization of drift tubes may be helpful
in reducing costs of manufacture of mobility spectrometers,
8
the
best route to low cost detectors could be a combination between
miniaturization and methods of mass production. A drift tube
that may be regarded as economical has been described and is a
planar micro-machined drift tube
9–11
with rectangular dimen-
sions of 25 mm long 3 0.5 mm deep 3 5 mm wide. This drift
tube was operated using a non-traditional approach with high
field asymmetric waveform dependent mobility methods.
12,13
Though the drift tube is small and ion losses to the walls might
be considered a complication, sub-nanogram detection limits
were obtained in the absence of ion shutters. Since these drift
tubes are manufactured using mass fabrication methods, the
most expensive part of the IMS analyzers, the drift tubes, now
can be made in mass with concomitant low costs. The size of the
analyzer and supporting electronics suggest that this design may
be attractive especially for field gas chromatographs and
process analyzers.
In a high field asymmetric waveform-operated drift tube, ions
are transported through the drift tube by a gas flow and electric
fields are applied perpendicular to the ion transport through a
planar drift region.
12,13
Unlike the behavior of ions in traditional
drift tubes, ion separation in this drift tube occurs according to
the ion mobility dependence on the electric field intensity. The
small size of the miniature mobility analyzer allows ion
residence times of about 1 ms and consequently the scan time
This journal is © The Royal Society of Chemistry 2002
466 Analyst, 2002, 127, 466–471 DOI: 10.1039/b111547m