IEEE TRANSACTIONS ON MEDICAL IMAGING, VOL. MI-4, NO. 3, SEPTEMBER 1985
A Method for Reconstructing Images from Data
Obtained with a Hexagonal Bar Positron Camera
GERD MUEHLLEHNER, MEMBER, IEEE, JOEL S. KARP, MEMBER, IEEE, AND ALBERT GUVENIS
Abstract-This paper describes the algorithms and procedures de-
veloped for reconstructing images using the hexagonal bar positron
camera. This camera has six continuous position-sensitive detectors
which are completely stationary, and it has some special software re-
quirements. In particular, spatial nonlinearities in the detectors must
be removed in software, the large but sparsely populated data matrix
must be reduced in size, and the gaps in the data from the intersections
of the detectors must be compensated for. These problems have been
investigated, and an appropriate algorithm for this system has been
implemented. The effectiveness of this algorithm was evaluated by re-
constructing both real and simulated data.
I. INTRODUCTION
T HE hexagonal bar positron camera comprises six po-
sition-sensitive detectors arranged in a hexagon to
form a single image plane. The detectors consist of 500-
mm long continuous bars of NaI(T1) to which ten photo-
multipliers are coupled. The position of a scintillation
event along the bar is determined through Anger-type pro-
cessing electronics [1], [2].
Since this design differs from other conventional posi-
tron tomographs which use arrays of discrete detectors, it
requires additional software which corrects for the various
imperfections associated with continuous position-sensi-
tive detectors, as well as an image reconstruction algo-
rithm which compensates for the gaps that occur between
neighboring detectors since the camera is completely sta-
tionary. The imperfections include the nonuniformity of
response along the detectors and positional distortions in
the measurements. Rebinning the data into a sinogram
prior to reconstruction introduces sampling patterns which
must be corrected for as well.
In this paper, we will first describe the methods used to
correct the raw data. The overall image reconstruction
software will then be described. Finally, the results ob-
tained from real and simulated objects will be reported.
II. DISTORTION REMOVAL
Spatial distortions are systematic errors in the position-
ing of scintillation events. Each event results in a distri-
bution of light at the plane of the photocathodes. The cen-
troid of this distribution, as measured by the phototubes,
is used to define the position of the event. The measured
position, however, is only an approximate linear function
Manuscript received January 21, 1985; revised April 24, 1985. This work
was supported by the Department of Energy under Contract DE-AC-
80EV 10402.
The authors are with the Department of Radiology, Hospital of the Uni-
versity of Pennsylvania, Philadelphia, PA 19104.
of the true position because the light distribution is sam-
pled only every 50 mm, and also because of effects near
the edges of the crystal. In addition, the differential and
integral nonlinearities of the analog-to-digital converters
contribute to the spatial distortions.
Since there is a one-to-one correspondence between the
measured and true positions, the spatial distortions can be
measured and compensated for [3]. This correction is
achieved in practice by collimating a line source placed at
the center of the patient aperture and using a lead ring
which has 90 equidistant slits (see Fig. 1). This type of
collimator allows narrow beams of gamma rays to hit the
detectors at known positions. Since the actual positions of
the coincident gamma rays on the detector surface are
known, the distortion factors can be computed by taking
the difference between the measured and the actual posi-
tions. We obtain 360 factors (one distortion factor per de-
gree) by rotating the lead ring four times in 10 increments.
This allows us to calculate the distortion factors at ap-
proximately 8-mm intervals along the detector. A com-
plete set of distortion factors is formed by linearly inter-
polating values between the measured points since the
distortions change smoothly and linearly over short dis-
tances. Fig. 2 shows the data accumulated for a detector
pair. Projection of the data onto either axis gives a spec-
trum of peaks for the corresponding detector. In subse-
quent collections (when the lead ring is removed, of
course), both position coordinates are digitized for each
coincident event, and the "true" location is obtained by
adding the distortion factor to the digitized coordinate.
The accuracy of the distortion removal scheme was
evaluated by placing a point source midway between two
opposing detectors. Since annihilation events are colinear,
the measured position of a coincident event on one detec-
tor is linearly dependent on the measured position on the
other detector. Without distortion removal, the average
deviation on the detector from this linear dependence is
2.5 mm, whereas after distortion removal, the average de-
viation is only 0.5 mm. This method of distortion removal
can correct the position of a scintillation only to an aver-
age position, independently of the depth of interaction in
the crystal. Particularly for gamma rays entering the crys-
tal obliquely, the uncertainty about the depth of interaction
causes a lateral offset which results in some loss of spatial
resolution.
Fig. 3 shows data from a number of point sources which
have been rebinned into polar coordinates (see Section III
for a description of rebinning) both without [Fig. 3(a) and
0278-0062/85/0900-0134$01.00
©
1985 IEEE
134