Journal of Chnical Forensic Medicine (1998) 5, 4 5 ~ 8
© APS/Harcourt Brace & Co. Ltd 1998
ORIGINAL COMMUNICATION
Breath testing in patients with respiratory disability
M. S. Odell*, C. E McDonald*, J. Farrar J. S. Natsis*, J. F. Pretto t
* Victorian Insfitute of Forensic Medicine, Southbank Victoria, Australia ~Austin and Repatriation Medical
Centre, Heidelberg Victoria, Australia ~ Victoria Police, Brunswick Victoria, Australia
SUMMARY. The presence of chronic respiratory disease is sometimes claimed as a defence by drivers who
refuse or fail a breath test. Trials have been conducted with two breath analysis instruments used in the state of
Victoria, Australia in order to determine the respiratory parameters compatible with a satisfactory test. Patients
with restrictive and obstructive conditions participated in the trial that was conducted in the respiratory unit of
a large teaching hospital. The results showed that the evidential breath analyzer currently used in Victoria was
able to accept breath samples from subjects at the limit of respiratory function likely to be encountered in drivers.
It was also found that a subject capable of providing a breath sample for the current screening device should also
be able to provide an evidential sample. The trials provided useful information in evaluating defences of inability
to provide a breath sample due to respiratory incapacity.
Journal of Clinical Forensic Medicine (1998) 5, 45-48
INTRODUCTION
In Victoria, breath testing for alcohol is carried out
under provisions of the Road Safety Act 1986.1 This
act provides for compulsory breath testing of drivers
under certain circumstances. While inability to
provide a suitable breath sample results in a required
blood sample, refusal to provide a breath sample is an
offence under the Act and carries a heavy penalty.
Notwithstanding the requirement for a blood
sample to be provided in the event of a failed breath
test, a number of drivers have attempted to use the
presence of chronic respiratory disease as a defence
against the charge of refusing a breath test. This trial
was carried out in an attempt to quantify the minimum
respiratory effort required for a satisfactory breath test
Dr Morris S. Odell BE(Hons), MBBS, FRACGP, DMJ, Forensic
Physician, Victorian Institute of ForensicMedicine, 57-83
Kavanagh Street, Southbank Victoria, 3006, Australia
Dr Christine E McDonald MBBS(Hons), FRACP, PhD,
Respiratory Physician,Austin and Repatriation Medical Centre,
John Farrar BSc(Hons),ForensicOffice1; Traffic Alcohol Section,
Victoria Police,20 Dawson Street, Brunswick Victoria 3056,
Australia, James S. Natsis, TraineeMedical Scientist,Austin and
Repatriation Medical Centre, Jeff J. Pretto BAppSe, Grad Dip
Biota Instr. Senior RespiratoryScientist, Austin and Repatriation
Medical Centre, StudleyRoad, Heidelberg Victoria 3084,
Australia.
Correspondence to." Dr Morris S. Odell
in order that such defences could be evaluated in the
light of formal respiratory function testing.
Two instruments are used by Victoria Police for
conducting breath tests: the Lion Alcolmeter SD2, a
portable hand-held device used for screening tests,
and the Drager Alcotest 7110 (Drager Australia,
3 Ferntree Place, Notting Hill 3168, Victoria,
Australia), which is a precision instrument used to
provide evidentiary print-outs accepted by the courts.
The Drager Alcotest 7110 which has been used in
Victoria since 1994, is specifically defined under the
Road Safety Act 1986' and is a substantially different
instrument to one with the same name used in other
jurisdictions. 2Each of the two instruments under eval-
uation contains sensors and circuitry for the purpose
of determining when an adequate breath sample has
been provided for analysis and will not produce a
reading unless pre-set internal criteria have been
satisfied.
The Lion Alcolmeter uses a simple timer to define
the sampling interval. The timer is started once the
breath flow exceeds a trigger level and runs for 2.7 s.
There is no defined criterion for flow rate during the
delay period other than to ensure that the dynamic
pressure generated by provision of the sample exceeds
the trigger level. Flow/pressure characteristics are
defined by disposable plastic tubular mouthpieces,
which are mass produced and have a high degree of
reproducibility.
45