Fresenius J Anal Chem (1994) 348:402-410 Freseniu$' Journal of
© Springer-Verlag 1994
A critical evaluation of the environmental scanning electron microscope
for the analysis of paint fragments in art conservation
N. W. Bower 1, D. C. Stulik 2, and E. Doehne 2
t Chemistry Department, Colorado College, Colorado Springs, CO 80903-3294, USA
2 The Getty Conservation Institute, 4503 Glencoe Avenue, Marina del Rey, CA 90292-6537, USA
Received March 29, 1993; revised July 5, 1993
Summary. Cross-sections from medieval paintings by Cenni
di Francesco and Dosso Dossi were analyzed for the inor-
ganic components as well as the binding media using an
environmental scanning electron microscope (E-SEM). The
advantages of this instrument compared to a normal SEM-
EDX are illustrated and a number of optimization studies
are reported. It was found that using a chamber gas pressure
of 1.5 kPa and a tungsten source instead of the usual LaB6
source with the 38 kPa pressure normally used for imaging
would significantly improve the x-ray analyses. Quantitative
analyses for most of the common medieval pigments are also
presented.
Introduction
The analysis of cross-sections of paint layers using
microchemical and instrumental methods is important for
the authentication and provenancing of works of art. The
information obtained from these studies are useful not only
for understanding the materials and techniques used by art-
ists, but are also needed for developing an appropriate proto-
col for the restoration or conservation of the work.
As early as /800 wet microchemical tests were used to
analyze artists' pigments in a study of medieval wall paint-
ings [1]. Since then a variety of techniques have been
employed, but light microscopy became the primary method
after Laurie [2] demonstrated its use for the study of paint
cross-sections at the beginning of this century. The appli-
cation of electron microprobe analysis (EMPA) for
microelemental analyses in conservation was demonstrated
as early as 1963 [3, 4]. Scanning electron microscopy coupled
with an energy dispersive detector (SEM-EDX) has ex-
panded this approach, and both methods are widely applied
in the field of art conservation today [5 - 7], although both
methods require that non-conductive paint specimens be
coated with a conductor and placed in a vacuum. Recent
developments in SEM technology [8, 9] have removed these
restrictions, opening the door to new applications and sim-
pler sample preparations.
Correspondence to: N. W. Bower
Theory
EMPA and SEM-EDX both use electron bombardment to
excite x-rays in the sample [10, 11] that are characteristic of
a sample's elemental composition. The degree of accuracy
for analyses made with standardless ZAF corrections is
approaching 5% for modern software packages from most
of the detector manufacturers. This is good enough to
correctly identify the common inorganic pigments used by
artists.
At atmospheric pressure an electron beam will be scat-
tered and absorbed and the lower energy x-rays emitted from
the sample may be attenuated. The heated filament normally
used to produce the electron beam will also be destroyed.
An additional problem is encountered with non-conductive
samples such as paint films. The bombarding electrons can
build up a local surface charge that is great enough to signifi-
cantly deflect the beam away from the point of interest. The
spreading of the incident electron beam from collisions with
gas molecules and from sample charging will negatively
impact the image quality of the electron micrograph and the
area being analyzed spectrally.
As a result of these problems, non-conductive specimens
are typically coated with a conductive layer of gold, osmium,
or carbon, The latter is best for x-ray analyses (though not
for image quality) as it does not produce spectral lines at
the energies typically detected with SEM-EDX instruments.
However, this coating may interfere with further analyses as
well as the image quality. For example, the identification of
the binding media components using FT-IR microscopy [12]
on the same sample would be difficult after the sample has
been coated. The IR spectra are usually obtained from an
adjacent cross-section obtained using a microtome. The IR
analysis may also be done on the same cross-section that is
used for the SEM-EDX analysis, if it is done before the
sample is coated. However, there are times when it is helpful
to reanalyze the same cross-section as more information is
obtained or as new questions arise.
The environmental scanning electron microscope (E-
SEM) circumvents some of these problems. It uses a moder-
ate pressure (0.75 to/50 kPa) of a gas such as water in the
sample chamber instead of coating the sample to remove the
excess charge [8, 9]. This is made feasible by a series of
apertures and vacuum pumps between the sample chamber
and the filament source.