MEASUREMENT OF ENTRANCE SKIN EXPOSURE (ESE)
VALUE WITH VARIOUS TYPES OF FILTER MATERIALS
ON MOBILE X-RAY MACHINE USING EXPOSURE
FACTOR OF CHEST
Muchamad Azam *, Evi Setiawati**, Fenia ***
*(Department of Physics, Diponegoro University, Semarang, ndonesia
azamfisika@gmail.com ,
**(Department of Physics, Diponegoro University, Semarang, Indonesia
evi_setiawati_msi@yahoo.com
***(Department of Physics, Diponegoro University, Semarang, Indonesia
feni@st.fisika.undip.ac.id
Manuscript History
Number: IJIRAE/RS/Vol.04/Issue10/OCAE10084
DOI: 10.26562/IJIRAE.2017.OCAE10084
Received: 22, September 2017
Final Correction: 30, September 2017
Final Accepted: 05, October 2017
Published: October 2017
Citation: Azam, M., Setiawati, E. & Fenia (2017). MEASUREMENT OF ENTRANCE SKIN EXPOSURE (ESE) VALUE
WITH VARIOUS TYPES OF FILTER MATERIALS ON MOBILE X-RAY MACHINE USING EXPOSURE FACTOR OF
CHEST. IJIRAE:: International Journal of Innovative Research in Advanced Engineering, Volume IV, doi:
10.26562/IJIRAE.2017.OCAE10084
Editor: Dr.A.Arul L.S, Chief Editor, IJIRAE, AM Publications, India
Copyright: ©2017 This is an open access article distributed under the terms of the Creative Commons Attribution
License, Which Permits unrestricted use, distribution, and reproduction in any medium, provided the original author
and source are credited.
Abstract— Measurement of Entrance Skin Exposure (ESE) value with a variety of types of material filter on
mobile X-ray machine has been done using exposure factor of chest examination. The type of filter used is
aluminum (Al), copper (Cu), galvalum (AlZnSi), and zinc (Zn). The tube voltage is varied at 70, 73, 77, and 81 kVp,
fixed current-flow 20 mAs, SSD distance 100 cm, and radiation area 10 cm x 10 cm. The result of measuring the
consistency of X-ray tube voltage has the biggest error of 1.52%. The exposure factor used in chest examination,
which is 80 kVp obtained by interpolation of data. The result of measuring ESE value with exposure factor of chest
(lateral) using filter 0.5 - 1.5 mmAlZnSi; 0.5 - 1.5 mmZn; and 0.5 - 1 mmCu are within tolerable limits. Similarly,
the ESE value measurement results with filters 1 - 3.5 mmAl; 0.5 mmCu; 0.5 - 1 mmAlZnSi; and 0.5 - 1 mmZn is
still within tolerable limits. The half ESE value of the 80 kVp exposure factor can be reached using a 3.2 mmAl
filter. This value is equivalent to using a 0.17 mmCu filter; 0.23 mmAlZnSi; or 0.23 mmZn.
Keywords— x-ray machine, ESE value, filter, HVL, multipurpose detector
I. INTRODUCTION
One radiology service is to use X-rays to support diagnostic and therapeutic processes. In X-ray imaging there is
an ESE or Entrance Skin Exposure value, ie the measured exposure value at the center of the X-ray axis which is
the center point of the region to be subjected to radiation. The ESE value is affected by the combination of targets
(anodes) and filters used in the X-ray tube. Filter is a material designed to absorb X-rays selectively.