XIX IMEKO World Congress
Fundamental and Applied Metrology
September 611, 2009, Lisbon, Portugal
A PVDF SENSOR WITH PRINTED ELECTRODES FOR NORMAL AND
SHEAR STRESS MEASUREMENTS ON SOLE
Satu Kärki
1
, Miika Kiiski
2
, Matti Mäntysalo
2
, Jukka Lekkala
1
1
Department of Automation Science and Engineering, Tampere University of Technology, Tampere, Finland,
satu.karki@tut.fi, jukka.lekkala@tut.fi
2
Department of Electronics, Tampere University of Technology, Tampere, Finland,
miika.kiiski@tut.fi, matti.mantysalo@tut.fi
Abstract We have tested a method of printing
electrodes on the unmetalled polyvinylidenefluoride (PVDF)
material to construct a matrix version of a sensor for normal
and shear stress measurements on sole. A commercial PVDF
material with silver ink metallization has previously been
used to manufacture a single sensor prototype. With the
metal-coated PVDF material, a matrix sensor is challenging
to construct; the metallization should be removed from the
certain areas of the material to form an electrode grid pattern
or a number of identical separate sensors should be cut off
from the material sheet. Hence, a new method is explored
here. The sensor is manufactured from unmetalled PVDF
material and an array of electrodes with desired size and
shape is printed on the material surface. This study
concentrates on the characteristics of single sensors
manufactured with this method. Based on the results, the
sensitivity seems to be decreased due to the thermal stress
caused by the electrode printing process. In the normal force
direction the sensor sensitivity was found to be about fifth
and in shear force directions about tenth of the
corresponding values measured with the sensor with
commercial electrodes. The sensitivity in this case, however,
is still adequate for stress measurements on sole.
Keywords: plantar stress, PVDF, printed electrodes.
1. INTRODUCTION
Measurement of plantar forces has indicated a
relationship between the excessive mechanical stress and
ulceration of the foot [1]. The pressure ulcers, also known as
pressure sores or decubitus ulcers, occur when the tissue is
compressed under pressure, e.g. due to a use of improper
footwear. An early identification of individuals at risk of
foot ulceration is one of the primary means to reduce the
incidence of ulceration [2]. Foot ulceration is a common
problem in people with diabetes mellitus and peripheral
neuropathy [3].
The mechanical stress on the plantar surface has two
components, pressure acting normal to the surface and shear
stress acting tangential to the surface [4]. The shear stress
can be further divided into anterior-posterior (AP) and
medial-lateral (ML) stresses [7]; the shear stress is a vector
addition of these two components [9]. The AP shear stress is
the horizontal component in the movement direction and the
ML shear stress the horizontal component perpendicular to
the movement direction [8].
Only normal stress is widely reported [5]. The main
reason is the lack of validated and commercially available
shear stress sensors [6]. During the last few decades,
however, a variety of methods has been developed for the
measurement of shear stress, see e.g. references [2-7] and
[9-10]. Shear stress on the skin interface has been shown to
increase blood flow occlusion in the deeper tissues,
generating stresses which are additional to those of normal
stress [5].
The aim of this study is to further develop a piezoelectric
polymer film sensor for plantar normal and shear stress
measurements, previously reported in reference [10]. At the
moment, only discrete measurements of plantar pressure can
be done with the sensor. The discrete sensor requires a
clinician to choose the appropriate location for optimal data
collection [11]. With an array of sensors, instead, plantar
pressure distribution over the entire plantar surface of the
foot could be assessed simultaneously [11]. Hence, we aim
to develop a matrix version of the sensor for on-floor and
also in-shoe plantar pressure measurements.
We have tested a method of printing electrodes directly
on the polyvinylidenefluoride (PVDF) material to
implement a matrix sensor. The single sensor prototype
reported in reference [10] was constructed manually from
commercial 28 ȝm thick PVDF material with silver ink
metallization, manufactured by Measurement Specialties
Inc. The sensor consists of four separate sensor elements
stacked together. Each piece of PVDF material was cut off
from the sheet and the wires were connected to the
electrodes with tin-plated copper tape. The construction of
an array of sensors with this method is time-consuming and
challenging: a number of identical separate sensors should
be cut off from the PVDF material sheet and manufactured
manually, or alternatively, the metallization should be
removed from the certain areas of the PVDF material sheet
to form an electrode grid pattern. Hence, here the sensor
elements are implemented from 28 ȝm thick unmetalled
PVDF material. The electrodes are printed on the material
sheet to form a grid pattern with desired array size and
shape. The thermal stress caused by the printing process
may affect the PVDF material properties, and thus, this
1765 ISBN 978-963-88410-0-1 © 2009 IMEKO