Clinical Feasibility of an Auto-Adjusting Bi-Level PAP Device
for the Treatment of Obstructive Sleep Apnea
Original Article / Özgün Makale
DOI: 10.4274/jtsm.015
Türk Uyku Tıbbi Dergisi 2014;3:66-70
Journal of Turkish Sleep Medicine 2014;3:66-70
66
Address for Correspondence/Yaz›flma Adresi: Nejat Altıntaş MD, Namık Kemal University Faculty of Medicine, Department of Pulmonary, Sleep and Critical Care,
Tekirdağ, Turkey E-mail: nejataltintas@gmail.com Received/Geliş Tarihi: 09.11.2014 Accepted/Kabul Tarihi: 11.11.2014
© Journal of Turkish Sleep Medicine, Published by Galenos Publishing. / © Türk Uyku Tıbbi Dergisi, Galenos Yayınevi tarafından basılmıştır.
1
Namık Kemal University Faculty of Medicine, Department of Pulmonary, Sleep and Critical Care, Tekirdağ, Turkey
2
Dışkapı Government Hospital, Clinic of Pulmonary, Sleep and Critical Care Medicine, Ankara, Turkey
3
University of Pittburgh Medical Center, Department of Pulmonary, Sleep and Critical Care Medicine, PA, USA
Nejat Altıntaş
1
, Burcu Oktay
2
, Hikmet Fırat
2
, Patrick Strollo
3
Obtrüktif Uyku Apne Tedavisinde Otomatik BPAP Cihazının Klinik Yararlılığı
Summary
Introduction: Bi-Level positive airway pressure (BPAP) is an effective
alternative to continuous positive airway pressure (CPAP) in patients
with obstructive sleep apnea (OSA) who could not tolerate CPAP.
An automatically titrating BPAP device has recently been developed,
BPAP Auto® with BiFlex™ (BPAPauto). The primary aim of this study
was to examine the performance of this new device during attended
polysomnography (PSG).
Materials and Methods: This was a prospective case series study.
Participants with OSA currently using CPAP or BPAP therapy were
recruited and undergo in-lab PSG study with BPAPauto.
Results: A total of 27 participants met the criteria, enrolled the study.
All participants received BPAPauto therapy during an attended PSG.
Sleep and respiratory data were examined. The mean apnea hypopnea
index was found 2.2±2.5 events/hour. SaO
2
(oxygen saturation) was
94.0%±1.8. The mean inspiratory positive airway pressure (IPAP) abolish
respiratory events was 14.1±3.4 cmH
2
O and that of expiratory positive
airway pressure (EPAP) was 10.7±3.9 cmH
2
O.
Conclusions: BPAPauto is able to establish an appropriate Bi-Level PAP and
control oxygen saturation without excessive disruption of sleep. Further
studies using randomized control design are needed to examine potential
roles and advantages of BPAPauto for treatment of OSA. (JTSM 2014;3:66-70)
Key Words: obstructive sleep apnea, bi-level continuous positive airway
pressure, continuous positive airway pressure
Özet
Giriş: CPAP’ı tolere edemeyen tıkayıcı uyku apne hastalarında BPAP,
CPAP’ın etkin bir alternatifidir. BPAP Auto® with BiFlex™ (BPAPauto) cihazı
son zamanlarda geliştirilen bir otomatik titrasyon özelliği olan BPAP’dır. Bu
çalışmanın amacı bu yeni cihazın etkinliğini laboratuvar ortamında yapılan
polisomnografi (PSG) ile değerlendirmektir.
Gereç ve Yöntem: Bu çalışma prospektif vaka-kontrol çalışmasıdır. Hale
hazırda BPAP veya CPAP cihazı kullanan obstrüktif uyku apneli hastalar
çalışmaya alındı ve BPAPauto cihazı ile laboratuar ortamında PSG yapıldı.
Bulgular: Çalışma kriterlerini karşılayan 27 hasta çalışmaya alındı. Tüm
çalışmaya alınan hastalara laboratuvar ortamında yapılan PSG eşliğinde
BPAPauto tedavisi verildi. Uyku ve solunum parametreleri incelendi.
Ortalama apne-hipoapne indeksi 2,2±2,5 olay/saat olarak bulundu.
SaO
2
(oksijen saturasyonu) %94,0±1,8 bulundu. Solunum olaylarının
düzeltmek için gereken ortalama inspiratuvar pozitif hava yolu basıncı
(IPAP) 14,1±3,4 cmH
2
O ve ortalama ekspiratuvar pozitif hava yolu basıncı
(EPAP) 10,7±3,9 cmH
2
O bulundu.
Sonuç: BPAPauto uyku kalitesini bozmaksızın, uygun BPAP basınçlarını
ayarlayabilmekte ve oksijen saturasyonunu kontrol altında tutabilmektedir.
Obstrüktif uyku apnenin tedavisinde BPAPauto’nun avantajlarının ve
potansiyel rollerinin incelene bilmesi için daha ileri randomize kontrollü
çalışmalara ihtiyaç vardır. (JTSM 2014;3:66-70)
Anahtar Kelimeler: Obstrüktif uyku apne, BPAP, CPAP