TUNER SYSTEM SIMULATION AND TESTS FOR THE 201-MHZ MICE
CAVITY
L. Somaschini, INFN-Pisa, Pisa, Italy
J. H. Gaynier, R. J. Pasquinelli, D. W. Peterson, R. Schultz, Fermilab, Batavia, IL, USA
Y. Torun, IIT, Chicago, USA
A. J. DeMello, A. R. Lambert, S. Virostek, LBNL, Berkeley, CA, USA
Abstract
The frequency of MICE cavities is controlled by
pneumatic tuners as their operation is impervious due to
large magnetic fields. The mechanical and RF transfer
functions of the tuner were simulated in ANSYS. The
first of these tuning systems was assembled and tested at
Fermilab. The mechanical response and the RF tuning
transfer function have been measured and compared with
simulation results. Finally the failure of different
actuators has been simulated and tested to predict the
operational limits of the tuner.
MICE CAVITY AND TUNER
Each acceleration module (RFCC) of the MICE
experiment [1] is composed of four normal-conducting
cavities made of copper with a nominal resonant
frequency of 201.25 MHz (fig. 1) [2]. The four cavities
will be mounted into a vacuum vessel in order to have the
same pressure on both sides and avoid deformation of the
cavity when pulling vacuum. The cavities will have to
work in a high magnetic field that can reach 3T on some
points of the cavities surface [3]: a tuning system able to
work in such an environment had to be designed. Each
cavity will be equipped with a pneumatic system of 6
tuners composed of an actuator and a fork. Each fork,
made of stainless steel, will be inside the vacuum vessel
and it will stretch and squeeze the cavity when operated
by the respective actuator. The actuators will be located
outside of the vacuum vessel and will share vacuum with
the vessel by means of bellows. All 6 actuators will be
connected in parallel and will be controlled by two
electronic valves, one responsible for pushing the actuator
shaft and the other one for pulling the shaft. The main aim
of the tuner will be to compensate for thermal drift of the
cavity and for structural differences between the 4 of
them, keeping all the cavities at the same resonant
frequency.
A test module with only one cavity has been built and
is about to be tested in the MuCool Test Area (MTA) at
FNAL [4].
TUNER CONTROLS
Pneumatic System
A control system for the MICE cavity tuner was
developed in order to operate the tuner both during test
and regular operation in the MTA. Compressed air is
provided by a compressor and is filtered to remove any
moisture. A relief valve prevents over-pressurization of
pneumatic system providing a relief at 120PSI. The
incoming air is routed to two proportional valves
manufactured by ProportionAir, respectively responsible
for the pressurization of the squeeze and stretch circuit of
the tuning system. Proportional valves are solenoidal
valves with a built in feedback loop. Once they are set to
reach a target output pressures, they regulate the airflow
to keep the output pressure constant, independently on the
input. Three additional electronic pressure gauges by
Omega with 4-20 mA current output provide a control
over input pressure, stretch and squeeze pressures. All
this pneumatic instrumentation is mounted on an
aluminum panel and every connection is realized with
copper plumbing. Two air lines connect the control panel,
placed away from the cavity, to the vacuum vessel. Here a
system of manual valves and manifold distributes the air
supply to each of the six actuators, allowing for an
independent exclusion of each actuator. Every connection
on this manifold panel is realized with copper while
radiation compatible sealant is used. Manual valves are all
made in stainless steel to be radiation compatible and
survive operations in the MTA.
Electronic System
The test system is equipped with dedicated control
electronics. The measured variables are pressure, resonant
frequency of the cavity and movement of actuator shafts.
The interface with the user is developed in LabView.
Here we have the possibility to set different target
pressures, to read digital pressure gauges, to monitor the
frequency read by a Network Analyzer and to monitor the
movement of each actuator shaft. The PC is directly
connected to the NWA and to an ADC by NI to acquire
the reading of 6 linear potentiometers used to monitor the
movement of each actuator shaft. At the same time the
Figure 1: The MTA cavity module and the tuning system.
5th International Particle Accelerator Conference IPAC2014, Dresden, Germany JACoW Publishing
ISBN: 978-3-95450-132-8 doi:10.18429/JACoW-IPAC2014-THPRI070
07 Accelerator Technology Main Systems
T06 Room Temperature RF
THPRI070
3927
Content from this work may be used under the terms of the CC BY 3.0 licence (© 2014). Any distribution of this work must maintain attribution to the author(s), title of the work, publisher, and DOI.