Research Article
Development of an Ex Vivo, Beating Heart Model for
CT Myocardial Perfusion
Gert Jan Pelgrim,
1
Marco Das,
2
Ulrike Haberland,
3
Cees Slump,
4
Astri Handayani,
1
Sjoerd van Tuijl,
5
Marco Stijnen,
5
Ernst Klotz,
3
Matthijs Oudkerk,
6
Joachim E. Wildberger,
2
and Rozemarijn Vliegenthart
1
1
University of Groningen, University Medical Center Groningen, Center for Medical Imaging-North East Netherlands,
Department of Radiology, Hanzeplein 1, 9713 GZ Groningen, Netherlands
2
Department of Radiology and Cardiovascular Research Institute Maastricht (CARIM), Maastricht University Medical Center,
Postbus 5800, 6202 AZ Maastricht, Netherlands
3
Siemens AG Healthcare, Forchheim, Germany
4
University of Twente, Drienerlolaan 5, 7522 NB Enschede, Netherlands
5
LifeTec Group BV, Den Dolech 2, 5612 AZ Eindhoven, Netherlands
6
University of Groningen, University Medical Center Groningen, Center for Medical Imaging-North East Netherlands,
Hanzeplein 1, 9713 GZ Groningen, Netherlands
Correspondence should be addressed to Rozemarijn Vliegenthart; r.vliegenthart@umcg.nl
Received 11 September 2014; Accepted 23 December 2014
Academic Editor: Marco Francone
Copyright © 2015 Gert Jan Pelgrim et al. Tis is an open access article distributed under the Creative Commons Attribution
License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly
cited.
Objective. To test the feasibility of a CT-compatible, ex vivo, perfused porcine heart model for myocardial perfusion CT imaging.
Methods. One porcine heart was perfused according to Langendorf. Dynamic perfusion scanning was performed with a second-
generation dual source CT scanner. Circulatory parameters like blood fow, aortic pressure, and heart rate were monitored
throughout the experiment. Stenosis was induced in the circumfex artery, controlled by a fractional fow reserve (FFR) pressure
wire. CT-derived myocardial perfusion parameters were analysed at FFR of 1 to 0.10/0.0. Results. CT images did not show major
artefacts due to interference of the model setup. Te pacemaker-induced heart rhythm was generally stable at 70 beats per minute.
During most of the experiment, blood fow was 0.9–1.0 L/min, and arterial pressure varied between 80 and 95 mm/Hg. Blood fow
decreased and arterial pressure increased by approximately 10% afer inducing a stenosis with FFR ≤ 0.50. Dynamic perfusion
scanning was possible across the range of stenosis grades. Perfusion parameters of circumfex-perfused myocardial segments were
afected at increasing stenosis grades. Conclusion. An adapted Langendorf porcine heart model is feasible in a CT environment. Tis
model provides control over physiological parameters and may allow in-depth validation of quantitative CT perfusion techniques.
1. Introduction
Computed tomography (CT) has become the premier non-
invasive imaging modality for the noninvasive evaluation
of the coronary arteries. For the functional assessment of
coronary artery disease (CAD), the sole diagnosis of coronary
luminal narrowing is ofen limited, especially in case of 30–70
percent (intermediate) grade stenosis [1]. Usually, additional
testing on the impact of stenosis on myocardial perfusion is
needed. To date, CT is not commonly used in daily clinical
practice worldwide, except for several leading clinics in CT
imaging. However, recent evidence suggests that state-of-the-
art CT scanners allow evaluation of myocardial blood supply,
on top of the interrogation of coronary morphology [2].
Tis includes quantifcation of myocardial perfusion using
dynamic perfusion techniques in second-generation dual
source CT (DSCT) scanning [3, 4]. Measurement of absolute
myocardial perfusion can enhance the diagnostic accuracy
for hemodynamically signifcant CAD, compared to visual
analysis of perfusion maps [4–6]. At present, only positron
emission tomography (PET) imaging is capable of true per-
fusion quantifcation [7]. Morton et al. recently demonstrated
Hindawi Publishing Corporation
BioMed Research International
Volume 2015, Article ID 412716, 8 pages
http://dx.doi.org/10.1155/2015/412716