Acoustic time-of-flight for proton range verification in water
Kevin C. Jones
Department of Radiation Oncology, University of Pennsylvania, Philadelphia, Pennsylvania 19104
François Vander Stappen
Ion Beam Applications SA, Louvain-la-Neuve 1348, Belgium
Chandra M. Sehgal
Department of Radiology, University of Pennsylvania, Philadelphia, Pennsylvania 19104
Stephen Avery
a)
Department of Radiation Oncology, University of Pennsylvania, Philadelphia, Pennsylvania 19104
(Received 29 March 2016; revised 27 July 2016; accepted for publication 3 August 2016;
published 25 August 2016)
Purpose: Measurement of the arrival times of thermoacoustic waves induced by pulsed proton dose
depositions (protoacoustics) may provide a proton range verification method. The goal of this study
is to characterize the required dose and protoacoustic proton range (distance) verification accuracy in
a homogeneous water medium at a hospital-based clinical cyclotron.
Methods: Gaussian-like proton pulses with 17 μs widths and instantaneous currents of 480 nA
(5.6 × 10
7
protons/pulse, 3.4 cGy/pulse at the Bragg peak) were generated by modulating the
cyclotron proton source with a function generator. After energy degradation, the 190 MeV proton
pulses irradiated a water phantom, and the generated protoacoustic emissions were measured by
a hydrophone. The detector position and proton pulse characteristics were varied. The experimental
results were compared to simulations. Different arrival time metrics derived from acoustic waveforms
were compared, and the accuracy of protoacoustic time-of-flight distance calculations was assessed.
Results: A 27 mPa noise level was observed in the treatment room during irradiation. At 5 cm
from the proton beam, an average maximum pressure of 5.2 mPa/1 × 10
7
protons (6.1 mGy at
the Bragg peak) was measured after irradiation with a proton pulse with 10%–90% rise time of
11 μs. Simulation and experiment arrival times agreed well, and the observed 2.4 μs delay between
simulation and experiment is attributed to the difference between the hydrophone’s acoustic and
geometric centers. Based on protoacoustic arrival times, the beam axis position was measured to
within ( x ,y ) = (−2.0, 0.5) ± 1 mm. After deconvolution of the exciting proton pulse, the protoacoustic
compression peak provided the most consistent measure of the distance to the Bragg peak, with an
error distribution with mean = −4.5 mm and standard deviation = 2.0 mm.
Conclusions: Based on water tank measurements at a clinical hospital-based cyclotron, protoa-
coustics is a potential method for measuring the beam’s position ( x and y within 2.0 mm) and
Bragg peak range (2.0 mm standard deviation), although range verification will require simulation
or experimental calibration to remove systematic error. Based on extrapolation, a protoacoustic
arrival time reproducibility of 1.5 μs (2.2 mm) is achievable with 2 Gy of total deposited dose.
Of the compared methods, deconvolution of the excitation proton pulse is the best technique for
extracting protoacoustic arrival times, particularly if there is variation in the proton pulse shape.
C
2016 American Association of Physicists in Medicine. [http://dx.doi.org/10.1118/1.4961120]
Key words: pulsed proton beam, range verification, acoustic pulses, ionoacoustics, protoacoustics
1. INTRODUCTION
Proton range uncertainty limits the benefits of proton therapy.
The proton’s unique dose deposition, which reaches its
maximum (the Bragg peak) at an energy-dependent depth
before abruptly falling off, allows for reduced entrance
and exit doses compared to photon therapy, which exhibits
monotonic decrease after the first few centimeters. Over-
shooting or under-shooting of the proton beam due to range
uncertainty, however, has large associated risks because of
the steep Bragg peak dose gradient. Range uncertainty is
primarily due to patient setup, changes in patient anatomy, CT
conversion, and treatment planning errors.
1
Improvements to
treatment planning
1
or CT conversion
2
may reduce the plan
range uncertainty, but they will not correct for setup error or
patient anatomy changes (unless imaging and planning are
performed before each fraction).
If possible, the ideal approach to reducing proton range
uncertainty is in vivo measurement of the proton penetration
depth.
3,4
Toward this goal, positron emission tomography
and prompt gamma monitoring are promising possibilities
that measure activation by protons, but they are limited by
the variable lifetimes of the emitting radioisotopes, energy
threshold for activation, and cost.
3
Another potential method,
5213 Med. Phys. 43 (9), September 2016 0094-2405/2016/43(9)/5213/12/$30.00 © 2016 Am. Assoc. Phys. Med. 5213