MOSFET dosimetry with high spatial resolution
in intense synchrotron-generated x-ray microbeams
E. A. Siegbahn,
a
E. Bräuer-Krisch, and A. Bravin
European Synchrotron Radiation Facility (ESRF), 6 Rue Jules Horowitz, 38043 Grenoble, France
H. Nettelbeck, M. L. F. Lerch, and A. B. Rosenfeld
Center for Medical Radiation Physics, University of Wollongong, Wollongong, New South Wales 2522,
Australia
Received 21 July 2008; revised 26 January 2009; accepted for publication 27 January 2009;
published 11 March 2009
Various dosimeters have been tested for assessing absorbed doses with microscopic spatial reso-
lution in targets irradiated by high-flux, synchrotron-generated, low-energy 30–300 keV x-ray
microbeams. A MOSFET detector has been used for this study since its radio sensitive element,
which is extraordinarily narrow 1 m, suits the main applications of interest, microbeam
radiation biology and microbeam radiation therapy MRT. In MRT, micrometer-wide, centimeter-
high, and vertically oriented swaths of tissue are irradiated by arrays of rectangular x-ray micro-
beams produced by a multislit collimator MSC. We used MOSFETs to measure the dose distri-
bution, produced by arrays of x-ray microbeams shaped by two different MSCs, in a tissue-
equivalent phantom. Doses were measured near the center of the arrays and maximum/minimum
peak/valley dose ratios PVDRs were calculated to determine how variations in heights and in
widths of the microbeams influenced this for the therapy, potentially important parameter. Monte
Carlo MC simulations of the absorbed dose distribution in the phantom were also performed. The
results show that when the heights of the irradiated swaths were below those applicable to clinical
therapy 1 mm the MC simulations produce estimates of PVDRs that are up to a factor of 3
higher than the measured values. For arrays of higher microbeams i.e., 25 m 1 cm instead of
25 500 m
2
, this difference between measured and simulated PVDRs becomes less than 50%.
Closer agreement was observed between the measured and simulated PVDRs for the Tecomet
®
MSC current collimator design than for the Archer MSC. Sources of discrepancies between
measured and simulated doses are discussed, of which the energy dependent response of the MOS-
FET was shown to be among the most important. © 2009 American Association of Physicists in
Medicine. DOI: 10.1118/1.3081934
Key words: x-ray, microbeam, radiation therapy, MOSFET, dosimetry, MRT
I. INTRODUCTION
The unique features of synchrotron radiation enable intense
monochromatic or polychromatic a spectrum of energies
x-ray beams to be used in biomedical experiments, some of
which are aimed toward radiotherapy.
1
This study focuses on
the dosimetry of microbeam radiation therapy MRT.
2,3
In
MRT, an array of rectangular x-ray beams, “microplanar”
beams, is used to irradiate selected biological targets. A typi-
cal microbeam array of area of 1 1 cm
2
contains 50
beams; each of these has a typical size of 1 cm 25 m
height width, separated by a center-to-center ctc spac-
ing of 200 m. The minimum dose in the central region
between two microbeams is called the “valley dose.” The
peak and valley dose ratios PVDRs in the irradiation field
are believed to be of importance for the therapeutic effect of
the treatment. Figure 1 presents a lateral dose profile in tis-
sue, typical for MRT, extracted along a line in the phantom
transverse the beam direction, where the peak and valley
doses which need to be measured are indicated. In the pre-
clinical in vivo MRT trials, these arrays irradiate the target
with either unidirectional or cross firing techniques.
4
In ra-
diotherapy, there are strict requirements on the precision with
which the delivered dose must be known.
5
Even if the accu-
racy requirements in standard hospital based radiation
therapy 3%Ref. 5 could be relaxed somewhat for MRT, it
would still be the physicist’s responsibility to ascertain the
dose as accurately as possible before therapy is implemented.
The x-ray source properties and the radiation interaction
probabilities in materials of interest can be used to simulate
the absorbed dose or it can be derived from measurement.
Accurate knowledge of the absorbed dose in tissue in MRT
pose unique challenges: the beam sizes used are small which
means that the detecting element must be small, ideally of
micron or submicron size; the x-ray energies used are in the
interval where many of the commonly used detectors have an
energy dependent response. Furthermore, the microbeam
dose deposition should also be measured at distances many
beam widths away from the direct irradiation field more
than 1000 m away from the field border which requires a
wide dynamic range in the dose detection.
Since the evolution of MRT, the production of planar
beams has been investigated with various collimator designs.
In the early MRT trials, a single slit was used to collimate the
1128 1128 Med. Phys. 36 „4…, April 2009 0094-2405/2009/36„4…/1128/10/$25.00 © 2009 Am. Assoc. Phys. Med.