Paper
MONTE CARLO SIMULATION OF AN ANTHROPOMETRIC
PHANTOM USED FOR CALIBRATING IN VIVO K-XRF
SPECTROSCOPY MEASUREMENTS OF STABLE LEAD IN BONE
Camille J. Lodwick* and Henry B. Spitz
†
Abstract—An anthropometric surrogate (phantom) of the hu-
man leg was defined in the constructs of the Monte Carlo N
Particle (MCNP) code to predict the response when used in
calibrating K x-ray fluorescence (K-XRF) spectrometry mea-
surements of stable lead in bone. The predicted response
compared favorably with measurements using the anthropo-
metric phantom containing a tibia with increasing stable lead
content. These benchmark measurements confirmed the valid-
ity of a modified MCNP code to accurately simulate K-XRF
spectrometry measurements of stable lead in bone. A second,
cylindrical leg phantom was simulated to determine whether
the shape of the calibration phantom is a significant factor in
evaluating K-XRF performance. Simulations of the cylindrical
and anthropometric calibration phantoms suggest that a cy-
lindrical calibration standard overestimates lead content of a
human leg up to 4%. A two-way analysis of variance deter-
mined that phantom shape is a statistically significant factor in
predicting the K-XRF response. These results suggest that an
anthropometric phantom provides a more accurate calibration
standard compared to the conventional cylindrical shape, and
that a cylindrical shape introduces a 4% positive bias in
measured lead values.
Health Phys. 95(6):744 –753; 2008
Key words: bones, human; Monte Carlo; physics, medical;
spectroscopy, gamma
INTRODUCTION
THE SKELETON of the human body is the major storage
location for calcium. Lead, as a congener of calcium, is
also stored in the skeleton. Persons who have been
chronically exposed to lead are at risk of serious health
effects. The risk is directly related to the amount of
exposure to lead and is especially high for young children
(Hu et al. 1998; Rabinowitz 1990). Although the con-
ventional method to detect exposure is to measure lead in
a sample of blood, direct measurement of stable lead in
bone is more indicative of cumulative, long-term expo-
sure than lead in blood (Ahlgren et al. 1976; Hu et al.
1989). In vivo x-ray fluorescence spectroscopy is a
non-invasive method for measuring stable lead in bone
that uses 88.034 keV photons emitted by a
109
Cd source
to excite the K shell electrons of stable lead atoms
deposited in bone. The intensity of the characteristic x
rays emitted as the electrons return to lower energy states
is directly related to the quantity of stable lead present in
bone (Todd and Chettle 1994; Somervaille et al. 1985).
Many studies have validated the accuracy of this method
by comparing in vivo K-XRF spectroscopy with chemi-
cal measurements of bones from cadavers of occupation-
ally exposed workers, revealing systematic bias was of
the order of 1–2 g Pb (g bone mineral)
-1
(Aro et al.
2000; Somervaille et al. 1986) or up to 5– 8 gg
-1
(Todd
et al. 2002). Being that the major interest in this diag-
nostic procedure remains in its reliability at concentra-
tions of lead less than 25 g Pb g
-1
and in vivo XRF
measurements are currently being used in epidemiologi-
cal studies of both workers and children, it is vital that
the method provides statistically meaningful results at
low concentrations.
Although the normalization technique does lessen
the variation observed between K-XRF measurements,
uncertainty exists. It has been reported that measurement
precision varies from person-to-person (inter-subject
variation) depending upon individual’s tissue thickness
and bone mineral mass despite the normalization process
(Aro et al. 1994). Additionally, uncertainty can be
correlated to variations in source-detector positioning
both during and between measurements (Spitz et al.
2000) as well as simple signal-to-noise counting statis-
tics. The objective of this work is to utilize simulations to
evaluate the effect of using an anthropometric calibration
phantom compared to the conventional cylindrical shape.
An advantage of Monte Carlo analysis is its ability to
* Oregon State University, Department of Nuclear Engineering
and Radiation Health Physics, 116 Radiation Center, Corvallis, OR
97331-5902;
†
University of Cincinnati, Department of Mechanical
Industrial and Nuclear Engineering, Cincinnati, OH 45221.
For correspondence contact: Camille Lodwick, Nuclear Engineer-
ing and Radiation Health Physics, 116 Radiation Center, Corvallis, OR
97331-5902, or email at camille.lodwick@oregonstate.edu.
(Manuscript accepted 21 May 2008)
0017-9078/08/0
Copyright © 2008 Health Physics Society
744