978-1-4673-9134-4/16/$31.00 ©2016 IEEE
Error Correction: a Proposal for a Standard
Harold Kirkham
1
, Artis Riepnieks
1
, Eddy So
2
and Jim McBride
3
1
Pacific Northwest National Laboratory
harold.kirkham@pnnl.gov
2
National Research Council of Canada
3
JMX Services, Inc
Abstract — Some of the errors in transducers such as
instrument transformers can be corrected as part of the digital
processing for the measurement. The instrument transformer can
be characterized in such a way that allows the Transducer
Electronic Data Sheet of IEEE Std 1451 to transfer the
information to the measurement system. A modification would
allow the measurement system to perform a high-quality self-
calibration whenever a transducer was replaced. That levies
requirements on the characterization accuracy of the instrument
transformer.
Index Terms — Transducer Electronic Data Sheet, transducer
error, self-calibration, error correction.
I. INTRODUCTION
Measurement connects the physical world to a conceptual
model: an equation. In a moving coil instrument, the “equals
sign” is manifested by the balance of the torque generated by
the current in the coil against the torque of a return spring. In a
digital instrument, the equation solving is more obvious. The
equation being solved is the measurand. It was shown in [1]
that the notion that measurement is equation-solving has many
useful consequences. The simplifications and approximations
of linguistic labeling are done away with: a concrete definition
is readily found for measurands that have been problematical
in the past—frequency, for example, when frequency is
changing. The residuals can be used to calculate a metric that
indicates the quality of the measurement.[2]
In this paper, we use the notion that the measurand is a
model, an equation, to examine high voltage measurement.
II. MEASUREMENT FRAMEWORK
The measurement framework, drawn as a block diagram,
establishes the relationships between parts of the measurement
process well known to metrologists. It can also be drawn to
show the transducers used, as in Figure 1. In the figure, the
solid arrows represent physical links, the open arrows
conceptual ones. Figure 1 shows a measurement system such
as a phasor measurement unit (PMU), in which the measuring
instrument is made as accurate as required, and calibrated
from its input terminals. In use, the measurement result will
inevitably contain the artifacts added by the transduction
system, typically using instrument transformers (IT).
Applied to high voltage measurement, an isolation
transformer, a CCVT or some active system based on field
measurement may be involved. The point is that the primary
quantity (the thing of concern to the application) is separated
from the measuring instrument. The realized quantity is
supposedly a scaled copy of the original, but the scaling will
be imperfect, to a greater or lesser degree.
III. CHANGING THE EQUATION
The artifacts of the transducer that cause differences from
perfect scaling are generally known as the cause of Type B
uncertainties. In particular, the scale factor could be off-
nominal, and there will be some frequency-response effects.
These uncertainties are generally characterized during
calibration tests.
For high voltage and current measurements, instrument
transformers (IT) are commonly used, with specified
“accuracy class.” Both IEC and IEEE specify the relationship
between ratio error and phase error graphically. The IEC
standard [3] shows the boundary of the permitted region by
rectangular boxes on a plane of ratio error vs phase error for
each IT accuracy class without restrictions on the range of
their load power factor. The IEEE standard [4] uses
parallelogram boxes to describe their accuracy class, with the
load power factor being limited within the range of 0.6 to 1.0.
There are proposals to change the IEEE standard to the
“square box” accuracy classes.
It is proposed that at least some of these artifacts can be
compensated for by changing the solution method in the
measurement algorithm. Calibration data for the transducer
can be added to the information available to the algorithm, and
Fig. 1. Measurement framework, with transducer included
• U.S. Government work not protected by U.S. copyright