© 2005 The Royal Microscopical Society
Journal of Microscopy, Vol. 217, Pt 3 March 2005, pp. 260–264
Received 16 July 2004; accepted 7 November 2004
Blackwell Publishing, Ltd.
SHORT TECHNICAL NOTE
Two-photon excitation fluorescence pH detection using
2,3-dicyanohydroquinone: a spectral ratiometric approach
P. D. JÖBSIS, C. A. COMBS* & R. S. BALABAN
Laboratory of Cardiac Energetics, *Light Microscopy Core, National Heart Lung and Blood Institute,
National Institutes of Health, Department of Health and Human Services, Bethesda, MD 20892,
U.S.A.
Key words. Cellular pH, dicyanohydroquinone, dihydroxyphthalonitrile, spectral
calibration, spectral imaging, two-photon.
Received 16 July 2004; accepted 7 November 2004
Summary
The use of 2,3-dicyanohydroquinone (DCHQ) as an emission
ratiometric probe of pH in vitro and in fibroblast cells was eval-
uated using two-photon excitation fluorescence microscopy
(TPEFM). In addition, methods for spectrally calibrating the
Zeiss LSM510 META spectroscopy system for TPEFM were
also developed. The emissions of both the acid and base forms
of DCHQ were detectable when using an 800-nm excitation
in TPEFM, thereby allowing ratiometric determination of pH.
These data suggest that, in contrast to most other emission
ratiometric probes, both acid and base forms of DCHQ have
similar two-photon cross-sectional areas at 800 nm. Acid
(maximum at ∼457 nm) and base (maximum at ∼489 nm)
DCHQ TPEFM emission spectra were similar to previously
reported one-photon excitation emission spectra. Calibration
curves for pH were successfully constructed using the ratio of
DCHQ emission difference maxima at 460 nm and 512 nm in
vitro and in cells. To our knowledge, DCHQ is currently the only
effective emission ratiometric pH indicator for two-photon
microscopy and may serve as a useful starting point for the
development of other TPEFM ratiometric dyes for quantitative
measurement of other cell parameters such as Ca
2+
, Mg
2+
or Na
+
.
Introduction
The use of spectral ratiometric approaches for monitoring
the cellular ionic milieu is highly desirable because effects of
bleaching, variations in excitation power or fluorophore con-
centration can be minimized. Ratiometric probes rely on the
spectral differences in the excitation or emission characteristics
of the probe as the ionic milieu shifts, providing an internal
reference. However, in two-photon excitation fluorescence
microscopy (TPEFM) (Denk et al., 1990) many of the ratiometric
probes that depend on either emission or excitation spectral
differences have not been found to be useful. The so-called
excitation ratiometric probes, such as Fura-2, do not have ade-
quate spectral excitation differences to differentiate the forms
of the dye due to the rather broad spectral density of TPEFM
excitation. The broad TPEFM excitation profile initially would
appear to be an advantage with emission ratio dyes because both
forms of the probes should be excited concomitantly. However,
with most probes the two-photon cross-sectional area of one form
of the dye dominates the emission properties, severely limiting
the ratiometric approach. The lack of adequate emission signal
from one form of a ratiometric probe has been demonstrated
by Baker et al. (2002) for the common pH probe SNARF and by
Szmacinski et al. (1996) for the calcium indicator INDO. In the
case of SNARF the base form dominates the emission charac-
teristics over the entire pH range. To our knowledge, no emission
ratiometric probes for pH are currently available for TPEFM.
The purpose of this study was to evaluate the two-photon
excitation characteristics of the ratiometric probe 2,3-
dicyanohydroquinone (DCHQ) (also known as 1,4 dihydroxy-
phthalonitrile; Kurtz & Balaban, 1986) as a TPEFM probe of pH.
Materials and methods
Calibration of the Zeiss META spectroscopy system in TPEFM
The META spectral imaging system (maximum resolution
10.7 nm) incorporated in the Zeiss LSM510 Axiovert 200 m
(Thornwood, NY, U.S.A.) microscope scan-head was used to
estimate the emission spectrum of DCHQ in TPEFM. Several
aspects of the manufacturer’s suggested methods for spectral
calibration of the system were found to be suboptimal, so a
customized method was adopted as outlined below. Because
no visible excitation light is present in TPEFM, we removed the
visible laser blocking pins in front of the META’s 32-channel
Correspondence to: Dr Paul Jobsis. Fax: +1 301 402 2389;
e-mail: Jobsisp@nhlbi.nih.gov