Copyright © Lippincott Williams & Wilkins. Unauthorized reproduction of this article is prohibited.
Influence of mechanical trauma of blood and hemolysis on
PFA-100 testing
Giuseppe Lippi
a
, Rossana Fontana
a
, Paola Avanzini
a
, Rosalia Aloe
a
,
Luigi Ippolito
a
, Franca Sandei
a
and Emmanuel J. Favaloro
b
Although the appropriate quality of samples is essential for
platelet function testing, information is lacking on
interference from mechanical trauma of blood and
hemolysis on PFA-100 analyzer. Citrated blood collected
from nine healthy volunteers was divided into three aliquots.
The first aliquot (‘A’) was processed without further
manipulation, whereas the second and third were subjected
to mechanical trauma by two (‘aliquot B’) or four passages
(‘aliquot C’) through a very fine needle (30 gauge) to
produce hemolysis and cell trauma mimicking poor sample
collection. Samples were tested on PFA-100 and Advia
2120, and plasma then separated and tested for lactate
dehydrogenase (LDH) and hemolysis index. Negligible
hemolysis was present in aliquot A (hemolysis index
0.2 W 0.1, cell-free hemoglobin 0–0.5 g/l), whereas an
increasing amount was present in aliquots B (hemolysis
index of 13.1 W 1.8, cell-free hemoglobin 6.0–6.5 g/l) and C
(hemolysis index 24.0 W 1.1, cell-free hemoglobin
11.5–12.0 g/l). Increases in LDH, and concomitant
reductions in platelet and red blood cell counts were
observed in aliquots B and C. In hemolyzed aliquots B, four
out of nine samples yielded ‘flow obstruction’ with both
PFA-100 agonist cartridges, whereas the closure times
were dramatically prolonged in the remaining five samples.
In hemolyzed aliquots C, flow obstruction was recorded in
six of nine samples for collagen and ADP and all samples for
collagen and epinephrine, whereas closure times of
collagen and ADP in the remaining three samples were
dramatically prolonged. Mechanical trauma of blood
causing hemolysis makes PFA-100 testing unreliable. When
flow obstructions are observed, the potential presence of
hemolysis should be investigated. Blood Coagul Fibrinolysis
23:82–86 ß 2011 Wolters Kluwer Health | Lippincott
Williams & Wilkins.
Blood Coagulation and Fibrinolysis 2012, 23:82–86
Keywords: hemolysis, hemolyzed specimens, interference, PFA-100 testing,
preanalytical variability
a
U.O. Diagnostica Ematochimica, Dipartimento di Patologia e Medicina di
Laboratorio, Azienda Ospedaliero-Universitaria di Parma, Parma, Italy and
b
Department of Haematology, Institute of Clinical Pathology and Medical
Research (ICPMR), Westmead Hospital, New South Wales, Australia
Correspondence to Professor Giuseppe Lippi, U.O. Diagnostica Ematochimica,
Azienda Ospedaliero-Universitaria di Parma, Via Gramsci 14, 43126 Parma, Italy
Tel: +39 0521 703050; fax: +39 0521 703197;
e-mail: glippi@ao.pr.it, ulippi@tin.it
Received 2 July 2011 Revised 15 August 2011
Accepted 27 August 2011
Introduction
The preanalytical phase is important, mandatory within
the total testing process, because it includes all the
manually intensive steps necessary to obtain samples
that are suitable for testing [1]. Nevertheless, flawed or
mishandled activities during the prelaboratory collection,
handling, and transportation of blood, as well as in the
laboratory preparation of test specimens might dramatic-
ally influence the reliability of a wide range of clinical
chemistry, immunochemistry as well as hematology test
results [1]. The quality of the blood sample is also
essential in the coagulation or hemostasis laboratory,
including that for platelet function testing, since unreli-
able results, potentially arising from ‘poor’-quality speci-
mens might lead to misdiagnosis of platelet function
disorders and von Willebrand disease (VWD), as well
as inappropriate perceptions and clinical responses
related to antiplatelet therapy [2–11].
Among the various preanalytical problems, spurious
hemolysis is known to reflect the most frequently
encountered preanalytical problem in specimens referred
for both clinical chemistry [12,13] and coagulation testing
[14]. In these settings, serum or plasma are the reference
sample matrixes, which are obtained after centrifugation
of whole anticoagulated blood. As such, the potential
presence of hemolysis can be reliably ascertained and the
degree of interference established. However, platelet
function testing is often performed on whole anticoagu-
lated blood [6,11,15], and it is thereby virtually imposs-
ible to establish whether a certain degree of hemolysis
has occurred or to what extent the poor quality of the
specimen might affect test results.
The PFA-100 analyzer is a reliable tool for screening
platelet dysfunction and/or VWD, which is also charac-
terized by acceptable sensitivity to different types of
antiplatelet drugs [7,16–18]. According to manufacturer’s
specifications, the leading quality requirements for pro-
cessing a specimen for platelet function on this compact
analyzer currently includes a visual assessment to ensure
adequate filling of the primary tube and the lack of visible
clots or aggregates. As such, hemolysis is not currently
listed among the potential ‘interferences’, nor has its
influence been assessed in the current scientific litera-
ture. Therefore, the aim of this study was to investigate
82 Technical report
0957-5235 ß 2011 Wolters Kluwer Health | Lippincott Williams & Wilkins DOI:10.1097/MBC.0b013e32834c6cb5