Combining Microfluidic Secretion Sampling with Small-Volume Proximity
Immunoassays: Application to Murine Islets and Adipocytes
Joonyul Kim*, Leah A. Godwin*, Kennon S. Deal*, Zac Keenum*, Desiree Wanders**, Robert L.
Judd**, and Christopher J. Easley*
* Chemistry and Biochemistry Department, Auburn University, 179 Chemistry Building, Auburn,
AL 36849-5312
** Anatomy, Physiology, and Pharmacology Department, Auburn University, 219 Greene Hall,
Auburn, AL 36849-5312
The microfluidic platform provides unique opportunities for manipulation and imaging of cellular
systems. Various approaches have been developed for culturing, imaging, and sampling from cells
in polydimethylsiloxane (PDMS) microdevices [1-3]. However, it is difficult to adapt generally-
applicable immunoassays to the micro- or nanoliter volumes sampled from microfluidic systems. As
such, micro-volume assays for specific proteins are typically specialized and lack flexibility for
application to other analytes.
Proximity ligation assays (PLA) are among the most sensitive, specific, and flexible protein assays
developed to date [4,5]. We have recently developed a modified version of these assays, termed
asymmetric proximity ligation assays (aPLA), where assay dynamic range was increased by 100-
fold, and sensitivity and limit of detection were also improved [6]. These assays are capable of
detecting a wide variety of protein analytes—theoretically any protein with two antibodies—at
attomole (or lower) detection limits from only one microliter of sample. We have developed aPLAs
for detection of thrombin [6], insulin (FIG. 1a), and adiponectin multimers (FIG. 1b). The dynamic
range of the insulin aPLA encompasses the entire range of multiple commercially-available enzyme-
linked immunosorbent assays (ELISAs), without requiring washing steps. Interestingly, we show
that our adiponectin aPLA is more sensitive toward high-molecular-weight multimers of the protein,
which have been shown to be the most biologically active forms. These aPLAs are ideal for
measurements on microliter-scale secretion samplings from cells in microfluidic systems.
We have also recently developed methods for secretion sampling from eight single pancreatic islets
of Langerhans in parallel on a microfluidic device (FIG. 2a). A key advantage offered by our
methodology is the passive nature of flow control, where only a single handheld syringe is required
to apply vacuum to the outlets of microfluidic channels. Flow rate is passively and accurately
controlled by precise patterning of microchannels to specify fluidic flow resistances. Sampling of
insulin secretion from many single islets was accomplished, and ELISAs were used to measure
insulin. Microfluidic sampling results from single islets were comparable to bulk, tube-based
measurements of islets in groups of ten (FIG. 2b), and single-islet volumes could also be acquired
using confocal reflectance microscopy (not shown). As shown in previous work [1], it should be
possible to modify the microfluidic system for passive sampling into aqueous-in-oil droplets, thereby
greatly improving temporal resolution on secretion measurements.
By combining homogeneous, small-volume protein assays such as aPLA with microfluidic
sampling, measurements can be expanded to a variety of cytokines or adipokines. For example,
insulin or glucagon from islets, as well as adiponectin or leptin secreted from adipocytes. We have
1224