153 Copyright © 2020, IGI Global. Copying or distributing in print or electronic forms without written permission of IGI Global is prohibited. Chapter 10 DOI: 10.4018/978-1-7998-1690-4.ch010 ABSTRACT The objective of project is to reduce the micromotion of novel implant under the static loads using func- tion of uniform design for FE analysis. Integrating the features of regular implant, a new implant model has been done. Micromotion of the novel implant was obtained using static structural FE analysis. Compared to the existing International team for implantology implants, the micromotion of the novel implant model was considerably decreased by static structural analysis. Six control factors were taken for achieving minimizes the micromotion of novel dental implant system. In the present work, uniform design technique was used to create a set of fnite element analysis simulation: according to the uni- form design method, all FE analysis simulation; compared to the original model, the micromotion is 0.01944mm and micromotion of improved design version is 0.01244mm. The improvement rate for the micromotion is 35.02%. Micromotion Analysis of a Dental Implant System R. Manimaran SRM Institute of Science and Technology, India Vamsi Krishna Dommeti https://orcid.org/0000-0003-0633-9825 SRM Institute of Science and Technology, India Emil Nutu University Politehnica of Bucharest, Romania Sandipan Roy SRM Institute of Science and Technology, India