Automating linear accelerator quality assurance
Tobias Eckhause
Department of Radiation Oncology, University of Michigan, Ann Arbor, Michigan 48109-5010
Hania Al-Hallaq
Department of Radiation Oncology and Cellular Oncology, The University of Chicago, Chicago, Illinois 60637
Timothy Ritter
Ann Arbor VA Medical Center, Ann Arbor, Michigan 48109
John DeMarco
Department of Radiation Oncology, Cedars-Sinai Medical Center, Los Angeles, California, 90048
Karl Farrey
Department of Radiation Oncology and Cellular Oncology, The University of Chicago, Chicago, Illinois 60637
Todd Pawlicki and Gwe-Ya Kim
UCSD Medical Center, La Jolla, California 92093
Richard Popple
Department of Radiation Oncology, University of Alabama Birmingham, Birmingham, Alabama 35249
Vijeshwar Sharma
Karmanos Cancer Institute, McLaren-Flint, Flint, Michigan 48532
Mario Perez
Royal North Shore Hospital, Sydney, NSW 2065, Australia
SungYong Park
Karmanos Cancer Institute, McLaren-Flint, Flint, Michigan 48532
Jeremy T. Booth
Royal North Shore Hospital, Sydney, NSW 2065, Australia
Ryan Thorwarth and Jean M. Moran
a)
Department of Radiation Oncology, University of Michigan, Ann Arbor, Michigan 48109-5010
(Received 17 May 2015; revised 5 September 2015; accepted for publication 9 September 2015;
published 25 September 2015)
Purpose: The purpose of this study was 2-fold. One purpose was to develop an automated, stream-
lined quality assurance (QA) program for use by multiple centers. The second purpose was to evaluate
machine performance over time for multiple centers using linear accelerator (Linac) log files and
electronic portal images. The authors sought to evaluate variations in Linac performance to establish
as a reference for other centers.
Methods: The authors developed analytical software tools for a QA program using both log files and
electronic portal imaging device (EPID) measurements. The first tool is a general analysis tool which
can read and visually represent data in the log file. This tool, which can be used to automatically
analyze patient treatment or QA log files, examines the files for Linac deviations which exceed
thresholds. The second set of tools consists of a test suite of QA fields, a standard phantom, and
software to collect information from the log files on deviations from the expected values. The test
suite was designed to focus on the mechanical tests of the Linac to include jaw, MLC, and collimator
positions during static, IMRT, and volumetric modulated arc therapy delivery. A consortium of eight
institutions delivered the test suite at monthly or weekly intervals on each Linac using a standard
phantom. The behavior of various components was analyzed for eight TrueBeam Linacs.
Results: For the EPID and trajectory log file analysis, all observed deviations which exceeded
established thresholds for Linac behavior resulted in a beam hold off. In the absence of an interlock-
triggering event, the maximum observed log file deviations between the expected and actual compo-
nent positions (such as MLC leaves) varied from less than 1% to 26% of published tolerance thresh-
olds. The maximum and standard deviations of the variations due to gantry sag, collimator angle, jaw
position, and MLC positions are presented. Gantry sag among Linacs was 0.336 ± 0.072 mm. The
standard deviation in MLC position, as determined by EPID measurements, across the consortium
was 0.33 mm for IMRT fields. With respect to the log files, the deviations between expected and
actual positions for parameters were small (<0.12 mm) for all Linacs. Considering both log files
6074 Med. Phys. 42 (10), October 2015 0094-2405/2015/42(10)/6074/10/$30.00 © 2015 Am. Assoc. Phys. Med. 6074