Impact of a new sampling buffer on faecal haemoglobin
stability in a colorectal cancer screening programme by the
faecal immunochemical test
Grazia Grazzini
a
, Leonardo Ventura
b
, Tiziana Rubeca
c
, Stefano Rapi
d
,
Filippo Cellai
c
, Pietro P. Di Dia
a
, Beatrice Mallardi
a
, Paola Mantellini
a
,
Marco Zappa
b
and Guido Castiglione
a
Haemoglobin (Hb) stability in faecal samples is an
important issue in colorectal cancer screening by the faecal
immunochemical test (FIT) for Hb. This study evaluated the
performance of the FIT-Hb (OC-Sensor Eiken) used in the
Florence screening programme by comparing two different
formulations of the buffer, both in an analytical and in a
clinical setting. In the laboratory simulation, six faecal pools
(three in each buffer type) were stored at different
temperatures and analysed eight times in 10 replicates over
21 days. In the clinical setting, 7695 screenees returned two
samples, using both the old and the new specimen
collection device (SCD). In the laboratory simulation, 5 days
from sample preparation with the buffer of the old SCD, the
Hb concentration decreased by 40% at room temperature
(25°C, range 22–28°C) and up to 60% at outside
temperature (29°C, range 16–39°C), whereas with the new
one, Hb concentration decreased by 10%. In the clinical
setting, a higher mean Hb concentration with the new SCD
compared with the old one was found (6.3 vs. 5.0 μg Hb/g
faeces, respectively, P < 0.001); no statistically significant
difference was found in the probability of having a positive
result in the two SCDs. Better Hb stability was observed
with the new buffer under laboratory conditions, but no
difference was found in the clinical performance. In our
study, only marginal advantages arise from the new buffer.
Improvements in sample stability represent a significant
target in the screening setting. European Journal of Cancer
Prevention 26:285–291 Copyright © 2017 Wolters Kluwer
Health, Inc. All rights reserved.
European Journal of Cancer Prevention 2017, 26:285–291
Keywords: cancer prevention, colorectal cancer screening,
faecal immunochemical test, sample stability
a
Screening Unit,
b
Clinical and Descriptive Epidemiology Unit,
c
Laboratory Unit,
Cancer Prevention and Research Institute (ISPO) Florence and
d
General
Laboratory, Laboratory Department, Careggi Hospital, Florence, Italy
Correspondence to Grazia Grazzini, MD, Screening Unit, Cancer Prevention and
Research Institute (ISPO), Via Cosimo il Vecchio 2, Florence 50139, Italy
Tel: +39 055 326 87279; fax: +39 055 326 97805;
e-mail: g.grazzini@ispo.toscana.it
Received 30 November 2015 Accepted 6 April 2016
Background
Screening for colorectal cancer (CRC) using guaiac faecal
occult blood tests (g-FOBT) has been shown to reduce
CRC-related mortality (Hewitson et al., 2008). Faecal
immunochemical tests (FIT) have gained international
acceptance as the best candidate to succeed g-FOBT
(Halloran et al., 2012). Recently, concerns have emerged
about the stability of FIT samples. A study by Van
Rossum et al. (2009) reported a significantly lower
adenoma detection rate (DR) in delayed returned
samples. Faecal haemoglobin (f-Hb) degradation seems
to occur more frequently in FIT samples with respect to
guaiac tests, probably because of the fact that
globin chains degrade more rapidly than haem. As FIT
analysis is based on antibodies specific for globin, false-
negative results are more likely to occur if samples are
exposed to globin degradation (Allison et al., 2014).
Moreover, the degradation of Hb, because of bacterial
enzymatic activity, may occur faster in samples collected
in buffer as used by most quantitative FIT methods
in comparison with the dry smears used for
g-FOBT cards.
A study carried out within our screening programme
showed a seasonal variation in positivity rates because of
the exposure of faecal samples to a high ambient tem-
perature (Grazzini et al., 2010). Another analysis (Van
Roon et al., 2012) showed that a prolonged sample return
time up to 10 days did not affect detection of colonic
lesions, although a seasonal variation in Hb concentration
was observed.
Recently, some changes in the specimen collection
device (SCD) used in the OC Sensor system have been
introduced by the manufacturer. Although overall chan-
ges are not specified, the main differences arise in the
composition of the buffer. This new formulation of the
buffer could make f-Hb less susceptible to influence by
timing or external physical factors. However, the che-
mical composition of the new buffer is not available and
no data are reported in the literature on the performance
of the new SCD.
The aim of this study was to compare the performance of
the new SCD with the one being used currently in the
Research paper 285
0959-8278 Copyright © 2017 Wolters Kluwer Health, Inc. All rights reserved. DOI: 10.1097/CEJ.0000000000000257
Copyright r 2017 Wolters Kluwer Health, Inc. All rights reserved.